WiseSpin CF-10 Personal Centrifuge Operation Manual (Version 1.4.2) written for Model CF-10
Contents Page 1. Safety Instructions........ 3 2. Before Use...... 5 3. Introduction... 6 4. Package Contents... 7 5. Appearance and Naming... 8 6. Controller Details... 9 7. Installation...... 10 8. Cautions in Use... 12 9. Operation... 14 10. Specifications... 18 11. Maintenance... 19 12. Certificate...... 20 WiseSpin CF-10 Personal Table-top Centrifuge 2/37
Safety Instructions 1 Please read this carefully for the safety before using the device. The indicated cautions are related to the safety and you should observe all safety and warning instructions. Symbols (Pictograms) and the meaning This symbol indicates caution (Including warning). There will be concrete descriptions. This symbol indicates inhibition regarding modification. This symbol indicates the risk of explosion or fire. This symbol indicates inhibition. There will be concrete description. This symbol indicates caution regarding heat. This symbol indicates the user must be kept away while operation. Please read following warning for safety and proper use. WARNING Do not use a power source other than which the unit is designed to operate on. Inappropriate power supply may result in damage to the equipment. Adhere to all electrical codes. Please do not touch with wet hands. It may be caused an electric shock. Do not handle the power cord harshly. Disconnection or short may cause if you bend or put a heavy thing on the device. When you take a plug out in a wall socket, turn the power off and then take the plug out carefully. WiseSpin CF-10 Personal Table-top Centrifuge 3/37
Insert the plug securely into a wall socket. Loose connections result in excessive heat generation to the device and may cause arcing to occur at the connection. Avoid using multi-point extension power cords. It may cause excessive heat generation and be the cause of a fire. If you extend the power cord, use the cord to have sufficient electric capacity. Be sure to operate the device only with the specified power supply. When you connect the product specified by 100V to 200V, it may be caused to over-heat, electric shock or a short circuit to each part. Do not connect wrong electricity since it is the cause of a fire. Disconnect the power supply plug if the unit is not going to be used for an extended period of time. Refer to the Warnings regarding Installation stated in the Section 7. Refer to the Warning and Cautions regarding use of the device stated in the Section 8. WiseSpin CF-10 Personal Table-top Centrifuge 4/37
Before Use 2 Thank you very much for purchase of WiseSpin CF-10 Centrifuge. This operation manual describes performance, proper usage, cautions and notices for use of the purchased product. Before use, please confirm this manual and use properly. Please read carefully and thoroughly all the safety instructions described in the Section 1 again. And also, please keep this manual. Modifications of exterior, interior, mechanism, electric/electronics or any other specifications could be made without notice to improve the quality and performance of the product. WiseSpin CF-10 Personal Table-top Centrifuge 5/37
Introduction 3 WiseSpin Digital Personal Table-top Centrifuges has very low noise level ( 57dB at Max. RPM) even though it has high performance (max. 13,500rpm (rcf: 12,225 x g)) and extremely compact with small footprint Key features are as follows; Unbelievably quiet operation with maintenance-free drive (Brushless DC motor) Very low noise level: 57dB at max. rpm High performance - Max. 13,500rpm (rcf: 12,225 x g) by digital feedback speed control Extremely compact design with small footprint Simplest user interface by Jog-Shuttle Switch - Patented user interface Exclusive Spin Down button: Two types of simple operation for repetitive use of a certain RPM & time. (Timer operating & Return operating) Acceleration and braking times lower than 15 sec (for max. Speed) Digital Back-Light LCD, with Power bar graph Digital timer function with continuous running mode (99hr 59min) Storage function for the set values of RPM and Timer Locking mode supported for experimental Safety Safety system for motor error and automatic stop when opening door Capacity: 12 x 1.5/2.0 ml tubes WiseSpin CF-10 Personal Table-top Centrifuge 6/37
Package Contents 4 CF-10 Personal table top Centrifuges.... [ 1 ] CFAR1512 Angle Rotor, 12 x 1.5/2.0 ml tubes... [ 1 ] Power Code..... [ 1 ] Operation Manual... [ 1 ] WiseSpin CF-10 Personal Table-top Centrifuge 7/37
Appearance and Naming 5 Air hole closed by a rubber cap in the factory setting. 3 2 1 5 4 6 1 Control Panel 2 Lid Clamp 3 Lid 4 Lid Shover 5 Lid Open Sensor 6 Rotor - CFAR1512 (12 x 1.5/2.0 ml tubes) WiseSpin CF-10 Personal Table-top Centrifuge 8/37
Controller Details 6 7 11 12 8 13 14 9 15 10 (NOTE) SV : Set Value, PV : Present Value 7 Operation Status 8 SV of rotation speed (set rpm : x 10) or Timer 9 PV of rotation speed (measured rpm : x 10) or Timer 10 Power bar graph indicating the power of the motor operating 11 ROTATION (operating) lamp : indication of rotation setting and status. 12 TIMER lamp : indication of Timer setting and status. 13 MODE button : Selection of main operation mode. (Set Rotation Speed mode, Set Timer mode and Safety Locking mode) 14 SPIN-DOWN button : Go to SPIN DOWN mode. 15 Jog-shuttle knob : Turn + Push - Setting the value of Rotation Speed or Timer by Turing. - Start and stop by Pushing WiseSpin CF-10 Personal Table-top Centrifuge 9/37
Installation 7 1. After receiving the unit and putting it on the specified place of use remove the wrapping. 2. Check the unit whether there are eventual transport damages or not. In case there are some transport damages found please inform the forwarder and your contractor. If there is no damage, please go to next point 3. 3. Push the Lid Clamp and open the lid. You can find the rotor, CFAR1512 (12 x 1.5/2.0 ml tubes) equipped in the centrifuging room. Please check carefully if there is anything wrong including loose fixing of the rotor or leaning of the rotor. In case there is any kind of damage found, please inform the forwarder and your contractor. If you cannot find anything wrong, please go to next point 4. 4. Connect the main power cord to the socket. WiseSpin CF-10 Personal Table-top Centrifuge 10/37
WARNING Place the device on a level, horizontal surface to support the capacity of the device without vibration. Do not install at a dangerous place. Do not operate the device where there are flammables. It could be caused serious accidents like explosion or fire. Do not operate the device outside. Avoid areas where the sun shines or rain directly on the device. It is very dangerous as cause over-heat or a short circuit. Do not tilt or lay down the device, and do not impact the body of the device. The device includes shock-sensitive motor and structures, which could be damaged by the tilt or impact. Do not install the device where it can get damp or dust. It may be caused overheating, a short circuit or a fire by damp or dust inside of the device. Do not use this device when temperature falls below 10 C and rise above 40 C. In door use only. Do not use the device at a place with high temperature. Do not operate the device near a heater, blowing an air-conditioner. It may be caused an accident by overheating of the motor. The device should be set at a well ventilated place. Be sure the holes on the side or bottom surface of the body are not blocked by some obstacles. Do not operate the device at a poor ventilated place or use closed. It may cause a fire or damage of the device as not to release heat or insufficient ventilation may result. Set the device a lightening place. When you set the device at a dark place, it may cause an unexpected accident by mistake in operation. If you want to run the machine for a long time, the rubber cap for the air hole in the lid should be removed before operation. If you run the machine for too long time with the air hole in the lid closed by rubber cap, it may cause damage to temperature sensitive samples. WiseSpin CF-10 Personal Table-top Centrifuge 11/37
Cautions in Use 8 WARNING Be careful to operate the device. You MUST read this manual carefully and completely before using. The device should only be operated by laboratory personnel especially trained for this purpose and familiar with all precautionary measures required for working in a laboratory. If unexpected sound, smell or smog is generated after connecting electric power, pull out the main plug and contact manufacturer or a supplier. Do not put any objects insides. When an object that is easy to be combustible is like a stone, a metal or paper goes insides, it may cause a fire, an electric shock or a break down. If do, pull out the mains plug from a socket and contact a manufacturer or a supplier. Do not disassemble or remodeling by yourself. It may cause damage to the designed safety. We do not have any duties for an accident, quality deterioration or a defect caused by a customer. All the mechanisms and electric/electronic devices in this machine are designed and developed by the manufacturer. The devices could be disassembled and serviced only by the engineers of manufacturer or by engineers approved by manufacturer. Do not touch the device with wet hand while operation. Do not turn on and off frequently. It could damage the electronic part. This device must be connected to an earthed power supply. WiseSpin CF-10 Personal Table-top Centrifuge 12/37
CAUTION Before every centrifugation operation, check that the rotor is completely and tightly fixed at the motor axis. Before every centrifugation operation, check that the lid is completely closed by the lid clamp. Do not open the door during rotation. Rotor will stop by safety system. But it needs 12 to 15 seconds. If you touch angle rotor before stop, may result serious injury. Basically, you MUST make the balance of angle rotor in loading sample to rotor. If you want to use the maximum rotor speed (13,500 rpm) for centrifugation, you SHOULD make the sample weight to fill up to two decimal places (i.e., 0.00g precision). If you don t keep this basic instruction, you can have a serious problem of the machine, and so it leads to an accident as well. Allow the rotation speed to slow zero before opening the lid when the centrifugation is finished. Please be careful that liquid should not penetrate into the part of control panel or back plane. Do not use flammable material. Flammable materials or Chemicals must not be used in order to avoid an explosion or a fire. Please avoid ultrasonic lamps. The unit could be cracked or faded. Disconnect the power supply plug if the unit is not going to be used for an extended period of time. WiseSpin CF-10 Personal Table-top Centrifuge 13/37
Operation 9 1. Power-On Turn on the power, and then get into its default mode. Install sample to rotor and close door completely. 2. Mode Selection You can change the mode by the combination of MODE button 13, SPIN button 14 and Jogshuttle knob 15 push. WiseSpin CF-10 Personal Table-top Centrifuge 14/37
SV 8 and PV 9 have different meanings for each operation mode. It is shown in the following <Table 1>. Setting Mode SV (Set Value) PV (Present Value) Control target Set Rotation Speed mode Set(Saved) Rotation Speed (rpm) Current Speed (rpm) Rotational Speed (rpm) Set Timer mode Set(Saved) Time Time remained Operation time Set SPIN Rotation Speed Mode Set(Saved) Rotation Speed (rpm) for SPIN operation Current Speed (rpm) for SPIN operation Rotation Speed (rpm) for SPIN operation Set SPIN Timer Mode Set(Saved) Time for SPIN operation Time remained for SPIN operation Operation time for SPIN operation Locking Mode Current Speed Current Time Display only <Table 1> the meaning of display by mode 3. Set Rotation Speed Mode After you turn on the power, the default mode of the unit is Set Rotation Speed mode. (NOTE) Please check carefully if the lid is completely closed before running. If the lid is not closed completely, the unit will not run by safety system. You can find zero or non-zero value of the SV rpm on LCD and ROTATION lamp 11 is blinking. If the SV is non-zero, it means that someone had saved a value previously by storage function. You can use this saved value or you can modify the value by turning the jogshuttle knob 15. Simply by clicking jog-shuttle, the unit begins to revolve at the set rpm. If the SV is 0, it means there is not the value saved previously. You can adjust the set rpm by turning jog-shuttle knob 15. RPM adjustment and Running: When you turn the jog-shuttle knob 15 for adjustment of rpm SV, the set value is changed. Click (push shortly) the jog-shuttle knob 15 to start rotational operation (centrifugation), then ROTATION lamp 11 is ON. Rotation rpm (PV) will goes up to set rpm (SV). Once the rpm PV reaches to rpm SV (within a few seconds), timer starts with its set WiseSpin CF-10 Personal Table-top Centrifuge 15/37
value and TIMER lamp 12 is ON. (rpm SV on the upper line of LCD changes to the value of remaining time). If you set timer SV to 0 before running the rotation, rotation will operate continuously without stopping unless you stop the rotation operation by clicking jog-shuttle knob in Set Rotation Speed mode. If you want to change the rpm SV, you have to stop the centrifugation by clicking jog-shuttle knob again. (Remaining time display changes to rpm SV. ROTATION lamp 11 is blinking and TIMER lamp 12 is OFF). Once stopping, you can adjust rpm again, and then click the jog-shuttle again to restart. In this case, timer will be reset to its SV, and the timer restarts from the beginning as well after reaching to target RPM. (NOTE) Control resolution is 100 rpm and maximum rpm is 13,500. Stop and Restart : If you click the jog-shuttle knob during rotation operation in Set Rotation Speed mode, the unit stops rotation and it completely stops after approximately 15 seconds (When you set SV to maximum specification 13,500 rpm). If you click again in this stop status, the unit will restart to operate. Saving current SV : You could save the current rpm SV for re-use in the next time by pushing jog-shuttle knob for one second. The unit stores the set value in nonvolatile memory, so you could re-use it afterwards. 4. Set Timer Mode You can use internal digital timer to make the operation preserved for a certain period of time. If you push MODE button 13 in Set Rotation Speed mode, you can enter into the Set Timer mode. You can find zero or non-zero value of the timer SV on LCD and TIMER lamp 12 is blinking. If the SV is non-zero, it means that someone had saved a value previously by storage function. You can use this saved value or you can modify the value by turning the jogshuttle knob 15. If the SV is 0, it means there is not value saved previously. You can adjust the timer SV by turning jog-shuttle knob 15. Timer adjustment and running Timer : When you turn the jog-shuttle knob 15 for adjustment WiseSpin CF-10 Personal Table-top Centrifuge 16/37
of timer SV, the set values is changed. Timer will automatically begin to run, at the moment that the rpm PV reaches to rpm SV by the Rotation operation. Saving current SV : You could save the current timer SV for the re-use in next time by pushing the jog-shuttle knob for one second. The unit stores the set value in nonvolatile memory, so you could re-use it afterwards. If you finish setting timer, push MODE button. you can enter into the Locking mode. (NOTE) Timer resolution is 1 second and maximum timer value is 99min 59sec. If you set timer to 0, rotation will be continued without stopping. 5. Locking Mode If you push MODE button in Set Timer mode, the unit enters into Safety Locking mode. In this mode, the PVs of rpm and timer are shown on LCD. And any action (turn or push) to jog-shuttle knob is disabled and pushing SPIN button is also disabled. So, the operator can prevent the unit from change of the operation by mistake and can monitor the operation safely. If you push the MODE button 13 again in this mode, the unit enters into the Set Rotation Speed mode. 6. SPIN mode This mode is very convenient for repetitive use of a certain combination of rotation speed & time. If you push SPIN button 14 in Set Rotation Speed mode or in Set Timer mode, the unit enters into the SPIN mode. By simply clicking the jog-shuttle knob at this moment, the unit starts operation promptly with the rotation speed and timer SVs saved previously. If you want to change the SVs of rotation speed and timer, push the jog-shuttle knob for one second. Then the unit enters into the Set SPIN Rotation Speed mode. Like in the Set Rotation Speed mode, you can change the rpm by turning jog-shuttle knob. After setup the rotation speed, push the SPIN button again (Save & go to next set up), then the current set rpm is saved and the unit enters into the Set SPIN Timer mode. Like in the Set Timer mode, you can change the timer SV by turning jog-shuttle knob. After setup the timer, push the SPIN button once again (Save & go to next set up), then the current timer SV is saved and the unit enters into the Set SPIN Rotation Speed mode again. When you finished setup of the rotation speed and timer for SPIN operation, push the jog-shuttle WiseSpin CF-10 Personal Table-top Centrifuge 17/37
knob for one second for returning to SPIN mode. By clicking the jog-shuttle knob, the unit starts operation promptly with the rotation speed and timer SVs saved just before. WiseSpin CF-10 Centrifuge offers two different types of SPIN mode operation. The one is Timer-based operation (with non-zero value of SPIN timer SV) and the other is One Shot operation (with zero-value of SPIN timer SV). If you set SPIN Timer to 0, the unit operates to reach SPIN rpm SV and then stops rotation immediately after the SPIN rpm PV reaches to SPIN rpm SV. This operation is very convenient for simple Spin-Down purpose. (It takes approximately 15 seconds to start, run and stop for spin-down operation, when your SPIN rpm SV is 6,000) If you want to return to normal operation mode, push the MODE button. 7. Power off After you complete to use the unit, you must turn off the power switch for your safety. (NOTE) If you open the lid during rotation, the unit will automatically stop the operation for user safety. Do not open the door during rotation!! If you don t keep this note, you can have a serious problem of the machine, and so it leads to an accident as well. WiseSpin CF-10 Personal Table-top Centrifuge 18/37
Specifications 10 CF-10 (with Angle rotor CFAR1512) Max. RPM Max. RCF Max. Capacity Control Display Timer Cooling Device Drive System Acceleration Time Braking Time Safety System Noise Level Dimension Weight Power Consumption Power 13,500 RPM 12,225 x g 1.5ml(2.0ml) x 12 Tubes Digital Feedback Control with Jog Shuttle Switch (Turn+Push) Digital Back-light LCD, with Power bar graph 99min. 59sec. (Can running continuous mode) Air Cooling Brushless DC motor, Direct drive 15 sec (for max. speed) 15 sec (for max. speed) Motor Error Detection, Auto Stop when opening door 57dB 170 W x 226 D x 183 H (mm) 2.3 kg 90W AC 110V, 50/60Hz or 230V, 50/60Hz WiseSpin CF-10 Personal Table-top Centrifuge 19/37
Maintenance 11 1. If you don t use the device for a long time, please plug out, dry up the machine and keep with packing. 2. When you need exchange of a component for repair, you MUST use an original component provided by manufacturer. 3. In the warranty period specified on the warranty sheet included in this operation manual, breakdown of the machine in spite of normal/proper use will be treated by manufacturer or by an official agent of manufacturer. 4. The manufacturer or its agent is not in charge of breakdown caused by abnormal use or by ignoring the warnings and cautions described in this manual. 5. Before cleaning the machine, please be sure to pull the power cord off. 6. For cleaning of the machine, we recommend you to use neutral detergent and soft cloth. Do not use coarse cloth, strong chemicals or organic solution. WiseSpin CF-10 Personal Table-top Centrifuge 20/37
CERTIFICATION of Quality Tested by Manufacturer and Free Warranty up to 2 Years for Laboratory Equipment Limited Warranty 1. This instrument has gone through our testing process about all functions before shipment. 2. This instrument is warranted for one year from purchasing date. 3. A breakdown caused by customer s misuse or natural disaster is excepted from this warranty. 4. This certification is only for international customers. Item Model No. Serial No. Date Customer WiseSpin Personal Table Top Centrifuges CF-10 DAIHAN Scientific, Co., Ltd. 24-4, Sangwolgok-Dong, Sungbuk-Ku Seoul, KOREA, 136-120 TEL: +82-2-967-5235, Ext. 129 FAX: +82-2-963-5231 Email: daihan@daihansci.co.kr www.daihan-sci.com f WiseSpin CF-10 Personal Table-top Centrifuge 21/37
WiseSpin CF-10 Personal Table-top Centrifuge 사용자설명서 (Version 1.4.2) written for Model CF-10 WiseSpin CF-10 Personal Table-top Centrifuge 22/37
목차 1. 사용하기전에... 23 2. 용도및특징... 24 3. 내용물.... 25 4. 외관및명칭 26 5. 동작 27 6. 주의사항. 33 7. 제품사양.. 35 8. 유지및관리 36 9. 품질보증 37 WiseSpin CF-10 Personal Table-top Centrifuge 23/37
사용하기전에 1 대한과학의 WiseSpin Digital Personal Table Top Centrifuges 를구입해주셔서감사합니다. 본설명서는구입하신제품의성능및사용법, 취급상의주의사항에대하여기술하고있습니다. 제품사용전에반드시주의깊게읽어주시기바랍니다. 사용자의주의가반드시필요한부분에는아래와같은경고표시가있으니각경고표시와그의미를숙지하시어사용시안전안전에관한주의사항에대해숙지하십시오. 경고주의사항을지키지않았을경우사용자가부상을입을수있음을알립니다. 주의이표시는주의사항을지키지않았을경우기기의고장의원인이됨을알립니다. 알림이표시는기능상주요한사항이나사용자가반드시숙지해야할사실을알립니다. 고온주의이표시는고온의히터를사용하므로화상의위험이있음을알립니다. 폭발위험이표시는휘발성, 폭발성화학물질을사용할경우폭발의위험이있음을알립니다. 접근금지기기의사용중에사용자의접근을제한함을알립니다. WiseSpin CF-10 Personal Table-top Centrifuge 24/37
용도및특징 2 WiseSpin Digital Personal Table Top Centrifuges 를구입해주셔서감사합니다. WiseSpin Digital Personal Table Top Centrifuges 는고성능 (Max. 13,500rpm (rcf: 12,225 x g)) 임에도불구하고동작중매우조용하며 ( 57dB at Max. RPM) 컴팩트한디자인을갖췄으며 다음과같은장점을가지고있습니다 : Brushless motor 사용으로조용하고진동이적은개인용테이블탑원심분리기 : 57dB at Max. RPM 최대 13,500rpm (rcf: 12,225 x g) 의고성능소형원심분리기 대한과학의특허! 조그셔틀조절기와디지털피드백기술적용으로편리한사용과정확한속도조절 Spin Down 전용버튼으로간단하게두가지의 Spin Down 용동작가능 ( 원하는 RPM에대한 Timer 동작과 Return 동작 ) 최고속도로의가속과최고속도에서의정지상태까지 15초이하의제어가능 백라이트기능을적용한시원한디지털 LCD 정보창 디지털타이머기능 (99hr 59min, 연속동작가능 ) RPM과 Timer 등의셋팅값저장기능. 기기동작시잘못된접촉등으로인한오동작방지기능 모터의오동작진단기능과안전한도어열림센싱기능으로사용자의안전제공 Capacity: 12 x 1.5/2.0 ml tubes WiseSpin CF-10 Personal Table-top Centrifuge 25/37
내용물 3 CF-10 Personal table top Centrifuges...... [ 1 ] CFAR1512 Angle Rotor, 12 x 1.5/2.0 ml tubes..... [ 1 ] Power Code... [ 1 ] User s Manual... [ 1 ] WiseSpin CF-10 Personal Table-top Centrifuge 26/37
외관 4 Air hole : 공장출하시이홀은고무캡으로막혀있습니다. 3 2 1 5 4 6 1 제어판 2 도어개폐기 3 도어 WiseSpin CF-10 Personal Table-top Centrifuge 27/37
4 도어지지대 5 도어열림센서 6 앵글로터 - CFAR1512 (12 x 1.5/2.0 ml tubes) 동작 5 Multifunctional Digital Feedback Controller 7 11 12 8 13 14 9 15 10 WiseSpin Digital Personal Table Top Centrifuges 의제어부. 7 동작상태표지 LCD 창 8 설정된값표시라인 (RPM 의경우 Set Value x 10) 9 현재동작값표시라인 (RPM 의경우 Measured Value x 10) 10 파워바그래프 : 현재동작하고있는모터의파워상태를그래프로표시 WiseSpin CF-10 Personal Table-top Centrifuge 28/37
11 회전 ( 원심분리 ) 동작상태를알려주는 LED. 12 타이머동작상태를알려주는 LED. 13 Mode 전환을위한버튼. 14 스핀다운 mode 전환을위한버튼 15 조그셔틀제어스위치 : Turning + Push 동작가이드 1. 전원동작 : 기기뒷면의주전원스위치를켜면, default mode 가됩니다. 원심분리할샘플을장착하고도어를꼭잠그십시오. 2. Mode 전환버튼 : Mode 전환스위치 13 를이용하여아래와같은 mode 들로전환. WiseSpin CF-10 Personal Table-top Centrifuge 29/37
아래표는모드별 LCD 상에표시되는내용에대해정리한것입니다 Setting Mode SV (Set Value) PV (Present Value) Control target Set Rotation Speed mode Set(Saved) Rotation Speed (rpm) Current Speed (rpm) Rotational Speed (rpm) Set Timer mode Set(Saved) Time Time remained Operation time Set SPIN Rotation Speed Mode Set(Saved) Rotation Speed (rpm) for SPIN operation Current Speed (rpm) for SPIN operation Rotation Speed (rpm) for SPIN operation Set SPIN Timer Mode Set(Saved) Time for SPIN operation Time remained for SPIN operation Operation time for SPIN operation Locking Mode Current Speed Current Time Display only 3. 회전동작 mode 전원을동작시킨후회전속도설정 mode 를선택하면 ROTATION LED 11 가 깜박거립니다. 그리고 LCD 상위표시창에는기존장비내에저장된설정된값이 있었다면설정값이표시되고, 설정된값이없다면 0 으로표시됩니다. 하위 표시창에는현재값들이표시되므로동작전에는 0 동작중에는현재의 회전속도 (rpm) 나온도가표시됩니다. i. 설정값이 0 이아니면, 이것은누군가이전에값을저장해두었다는 의미이며, 이저장값으로장비를동작시키려면조그셔틀스위치를살짝 클릭해주면됩니다. 또한조그셔틀스위치를돌려설정값을조정한후 조그셔틀을클릭하면조정된값으로장비가동작합니다. WiseSpin CF-10 Personal Table-top Centrifuge 30/37
ii. 설정값이 0 이라면, 이것은이전에아무값도저장되어있지않았다는 의미이며, 이상태에서는조그셔틀을돌려설정값변경이가능합니다. RPM 조정 : 조그셔틀스위치를돌려원하는속도를설정후조그셔틀스위치를 클릭하면회전 - 원심분리동작을시작하며 ROTATOIN ( 회전동작 ) LED 11 가 켜집니다. 이어곧목적 rpm 에도달하게되면이미저장된타이머가자동으로 동작을하며 TIMER LED 12 가켜집니다 (LCD 창의현재 RPM 표시가남은타이머 값으로변경 ). 만약타이머값을 0 으로설정했었다면회전동작은멈춤을 조작하기전까지영구히동작합니다. 만약동작중에회전속도를변경하여다시동작을시키고싶다면조그셔틀스위치를 눌러동작을먼저멈춘후에속도설정을다시할수있습니다. 이때 TIMER LED 12 는꺼지고 ROTATION LED 11 가다시깜박거리며동작했던타이머또한초기화 ( 설정값으로복귀 ) 됩니다. 재설정을마치고다시회전동작을시작하면목적 RPM 도달후타이머또한설정값으로부터다시동작합니다. (Control resolution: 100 rpm for speed) 정지와재시작 : 동작시키기전에도어가잠겼는지다시한번확인하시기 바랍니다. 만약도어가완전히잠기지않았다면기기는안전장치에의해 동작하지않습니다. 장비가회전동작모드에서회전동작을할때조그셔틀을 클릭하면회전을멈춥니다 ( 최고속도에서완전정지까지 15 초이하소요 ). WiseSpin CF-10 Personal Table-top Centrifuge 31/37
정지상태에서다시조그셔틀을클릭해주면다시회전을시작합니다. ( 정지상태에서최대속도까지 15 초이하소요 ). 현재설정값저장 : 조그셔틀스위치를 1 초동안누름으로써현재설정된회전속도 (rpm 값 ) 가장비내부의메모리에저장됩니다. 이값은전원을껐다가다시켠 다음에도계속장비에저장되어있어또다른조절없이편리하게이용하실수 있습니다 4. 타이머 mode 타이머 mode 는원하는시간 ( 설정시간 ) 동안만동작을하게하고설정시간 이후에는동작을멈추게하는기능을설정하는 mode 입니다. 타이머모드가 선택되면 TIMER LED 12 가깜박거립니다. LCD 상위표시창에는만일기존장비 내에저장된설정된값 ( 설정시간 ) 이있다면설정값이표시되고, 설정된값이 없다면 0 이표시됩니다. 타이머설정 / 작동 : 타이머를설정하기위해서 mode 선택버튼을눌러타이머설정 mode 로들어갑니다. 조그셔틀을돌려원하는설정시간을맞추면됩니다. ( 조그셔틀 클릭필요없음 ) 타이머는회전동작을시작했을때목적 RPM 에다다르면자동으로 동작을시작합니다. 다른 mode 에서타이머 mode 에다시들어가면설정해놓은 시간과현재잔여시간을볼수있습니다. WiseSpin CF-10 Personal Table-top Centrifuge 32/37
타이머설정은 1 초간격으로설정이가능하며최대입력값은 99 분 59 초입니다. 만약타이머설정을 0 으로하면회전동작은멈춤동작을하기전까지영구히 계속됩니다. 설정시간저장 : 마찬가지로타이머모드에서조그셔틀스위치를 1 초동안누르면 현재설정시간이장비내부의메모리에저장됩니다. 이값은전원을내렸다가켠 다음에도계속장비에저장되어있습니다. 5. Locking mode RPM 과타이머설정및동작시작후, mode 전환버튼 13 클릭하여장비를 Locking 모드상태로설정할수있습니다. Locking 모드에서장비는현재측정 rpm 과잔여 시간을표시해주며, 모드전환버튼을제외한조그셔틀스위치로부터의어떠한 입력도무시합니다. Locking 모드에서벗어나고자하면 mode 전환버튼을눌러주면 됩니다. 6. Spin Down mode Spin Down mode 는자주사용하는스핀다운기능을별도의버튼으로제공하여 사용상의편의를제공하고있습니다. 간단한침전을위해스핀다운기능을 사용하려면 SPIN button 14 을누르십시오. 스핀다운동작을위해조그셔틀스위치를클릭하면이전에저장한스핀다운 WiseSpin CF-10 Personal Table-top Centrifuge 33/37
회전속도와시간으로동작을시작합니다. 스핀다운동작의회전속도와동작시간을조절하려면조그셔틀스위치를 1 초간 눌러주십시오. 조그셔틀스위치를이용해 SPIN DOWN 동작을위한회전속도를 설정할수있습니다. 회전속도셋팅이끝난후에 SPIN button 14 ( 저장및다음단계이동 ) 을누르면 회전속도값이저장이되고타이머조절모드로들어갑니다. 조그셔틀스위치를이용해타이머값을설정한후, SPIN button 14 ( 저장및 다음단계이동 ) 을다시한번누르면타이머값이저장되고다시회전속도조절 모드가표시됩니다. 모든셋팅이완료가되고마치기위해선조그셔틀스위치를 다시 1 초간눌러줍니다. 스핀다운모드로다시돌아가게되며조그셔틀을누르면 ( 클릭 ) 조절한값으로 스핀다운동작이가능합니다. WiseSpin CF-10 원심분리기는두가지의스핀다운동작을지원합니다. 한가지는 위에설명한바와같이설정한 RPM 으로설정한시간만큼동작하는 Timer-based 동작이며다른한가지는설정한 RPM 까지도달후바로회전을멈추는동작을하는 One Shot 동작입니다. 이는 SPIN 모드타이머설정에서타이머값을 0 으로설정후동작을시키면이미 설정된 RPM 까지도달후바로멈춤동작을실행하는것으로간단한스핀다운 WiseSpin CF-10 Personal Table-top Centrifuge 34/37
동작시매우유용합니다. (6,000 RPM 설정시대략총 15 초간가속과감속동작을 하게됨.) 일반사용 mode 로돌아가기위해서는 mode 전환버튼를다시눌러주면됩니다. 7. 설정값의저장 본장비는각각의모드에서설정된값을장비의내부메모리에저장할수있는 기능을포함하고있습니다. 이는자주사용하는설정값들을전원을 OFF 시키더라도 다시전원을 ON 시켰을때별도의설정없이편리하게장비를사용할수있습니다. 각모드 ( 잠금모드를제외한모든모드 ) 에서조그셔틀스위치를이용하여원하는 설정값을설정합니다. 조그셔틀스위치를 1 초이상클릭하면 LCD 우측상단의설정값이깜박거리면서현재 설정값을저장합니다. 8. 전원끄기 안전을위해서모든사용을마친후장비의메인스위치를반드시꺼주시기 바랍니다. 원심분리기동작중임의로도어를열거나기타다른외부요인으로인해도어가열릴 경우 WiseSpin Digital Personal Table Top Centrifuges 는도어센서가장착되어있어 자동적으로회전동작을멈추게됩니다. 원심분리동작중에는절대로문을열지않도록 WiseSpin CF-10 Personal Table-top Centrifuge 35/37
하시기바랍니다. 만약동작중문을열경우기기에심각한손상을초래할수있을뿐만 아니라사용자의안전에도치명적인사고를일으킬수있습니다. WiseSpin CF-10 Personal Table-top Centrifuge 36/37
주의사항 6 본장비는꼭일반적으로사용되는적절한전원을사용하여주십시오. (AC 100, 50/60Hz or AC 230V, 50/60Hz, Single phase) 반드시허용된규정부하이 하에서만사용하십시오 본장비를사용하시기전에는꼭사용자매뉴얼을숙지하여주시고실험실의 매뉴얼을숙지하여인가된사람외에사용을금해주시기바랍니다. 사용중평소와다른소음이나동작징후를발견시에는즉시기기의전원을 끄시고전원선을차단시킨후기기를공급한곳이나 대한과학의기술지원 팀으로문의하시기바랍니다. 본장비동작전에 rotor 가제대로장착되어있는지확인하시기바랍니다. 또 한동작중에도어를열지마십시오. 안전장치에의해 rotor 는곧 (15 초이 내 ) 멈추지만그전에빠르게회전하고있는 rotor 를손으로만질경우심각한 상처를입을수있습니다. WiseSpin CF-10 Personal Table-top Centrifuge 37/37
젖은손으로기기를동작시키거나전원스위치를무리하게반복적으로동작하 지마시기바랍니다. 반드시안정된장소에서사용하여주시고기기를기울이거나하지마십시오. 휘발성, 가연성, 폭발성화학물질을사용할경우폭발의위험이있으니절대 로사용하지마시기바랍니다. 비인가된사람에의해장비를분해하거나변형시키지마십시오. 심각한전기 정문제를야기시킬수있으며 대한과학의보증서비스를받을수없습니다. 원심분리기사용시 rotor 에장착하는샘플은위치와무게가균형을이뤄야 합니다. 최고속도 13,500rpm 동작시최소한샘플의무게는소수둘째자리 gram 까지맞춰줄것을권장합니다 (0.0X g). 이주의사항을지키지않을시기기에심각한손상을끼칠수있으며이로인 해심각한사고를야기시킬수있습니다. 햇빛이지속적으로비추는곳이나비를맞을수있는곳에서의사용을금합니 다. 이는기기의과열이나전기적결함을일으킬수있습니다. ( 실외사용금지 ) WiseSpin CF-10 Personal Table-top Centrifuge 38/37
가열기기등이있는곳이나뜨거운바람이부는환경을피해주시기바랍니다. 이는기기의모터등에과열을일으켜손상을입힐수있습니다. 장시간연속적인사용을피해주시기바랍니다. 도어상부 Air hole 의고무캡 을제거한다해도장시간연속적으로사용하면 rotor 의온도가상온이상으로 올라갈수있으며이는온도에민감한샘플의경우실험결과에영향을미칠 수있습니다. 제어판넬과후면부에실험에사용되는용액이흘러내리지않도록주의하며 장비의플랫폼을때리거나큰힘으로누르지마십시오. 반드시실내에서동작하여주시고, 10 ~ 40 사이의온도를유지하여주십 시오 제품적절한수명을위해깨끗한사용환경을유지하여주십시오. 먼지가많은 곳에서의사용은환기를저해하여기기의오동작과과열을초래할수있습니 다 이제품에사용된모드회로및전기적부품들은 대한과학의기술적자 WiseSpin CF-10 Personal Table-top Centrifuge 39/37
산입니다. 절대 대한과학의서비스팀기술자나 대한과학의교육을통해 인정한기술자가아닌비인가자에게수리를맡기지마시기바랍니다. WiseSpin CF-10 Personal Table-top Centrifuge 40/37
제품사양 7 CF-10 (with Angle rotor CFAR1512) Max. RPM Max. RCF Max. Capacity Control Display Timer Cooling Device Drive System Acceleration Time Braking Time Safety System Noise Level Dimension Weight Power Consumption Power 13,500 RPM 12,225 x g 1.5ml(2.0ml) x 12 Tubes Digital Feedback Control with Jog Shuttle Switch (Turn+Push) Digital Back-light LCD, with Power bar graph 99min. 59sec. (Can running continuous mode) Air Cooling Brushless DC motor, Direct drive 15 sec (for max. speed) 15 sec (for max. speed) Motor Error Detection, Auto Stop when opening door 57dB 170 W x 226 D x 183 H (mm) 2.3 kg 90W AC 110V, 50/60Hz or 230V, 50/60Hz WiseSpin CF-10 Personal Table-top Centrifuge 41/37
유지및관리 8 1. 장기간사용하지않을시전원플러그를빼고건조시킨후포장하여보관합 니다. 2. 사용중고장으로인하여부품교환이필요한경우당사에서제공하는순정 부품만을사용하여야합니다. 3. 정상적인한계를벗어난사용으로인한고장은유지보수할수없습니다. 4. 제품을청소할시에는전원플러그를빼낸후부드러운수건으로닦으시고, 지워지지않는부분은어는점이낮은용제를수건에묻혀오물부분만닦으 십시오. WiseSpin CF-10 Personal Table-top Centrifuge 42/37
품질보증서 (CERTIFICATION) 이보증서의소정항목에기재되지않은사항이없는지확인하여주십시오. 이보증서의기재내용에따라보증하여드립니다. 이보증서는대한민국국내고객용이며아래보증내용을참조하십시오. 사용중고장이발생하였을경우구입처로서비스를요청하십시오. 이보증서는재발행되지않으니잘보관하여주십시오. 제품명 모델명 WiseSpin Personal Table Top Centrifuges CF-10 제조번호 보증기간 구입일로부터 12 개월 구입일자년월일 구입대리점 보증내용 1. 본제품은보증기간동안소비자의과실이아닌제품의하자로인한고장의경우에는무상으로수리또는동종의제품으로교환해드리겠습니다. 2. 품질보증기간은구입일로부터 1년 (12개월) 입니다. 유상서비스 다음과같은경우에는보증기간내에라도별도의서비스요금을청구합니다. 1. 품질보증기간 1년을경과한경우. 2. 제품을임의로개조또는수리한경우. 3. 사용자의취급부주의로인한고장인경우. 4. 천재지변 ( 화재, 염해, 가스해, 수해, 기타 ) 에의한고장인경우. ( 주 ) 대한과학 WiseSpin CF-10 Personal Table-top Centrifuge 43/37