특징 Reference Guides Class II Biological Safety Cabinet SterilGARD e3 SterilchemGARD III Advance Class III Biological Safety Cabinet IsoGARD Pharmacy Barrier Isolators ChemoSHIELD[negative pressure]glovebox Laboratory Animal Research Product AniGARD II Animal Transfer Station AniGARD VF Animal Transfer Station 241
Reference Guides Biological SafetyCabinet Classification Specifications New NSF Classification, Adopted 2002 Class ll, Type A1 Class ll, Type A2 General Description * Chamber air의 70% recirculation, 30% exhaust * 75 FPM intake * Contaminated air가 흐르는 cabinet 내부의 공기통로가 positive pressure으로 되어 있어 오염된 공기가 누출될 위험이 있음 * Chamber air의 70% recirculation, 30% exhaust * 100 FPM intake * Contaminated air가 흐르는 cabinet 내부 공기통로의 positive pressure는 negative pressure로 둘러싸여 있어 내부 공기가 누출되지 않음 * Chamber 내부에 vapour나 gas 발생 시 배출이 원활히 되지 않고 누적됨 Class ll, Type B1 * Chamber air의 40% recirculation, 60% exhaust이 되므로, 60%가 빠져나가려면 Hard connection duct 설치 시 출구쪽에 blower가 있어야 함. * 100 FPM intake * Contaminated air가 흐르는 cabinet 내부 공기통로의 positive pressure는 negative pressure으로 둘러싸여 있어 내부 공기가 누출되지 않음 Class ll, Type B2 * Chamber air가 100% exhaust * Hard exhaust connection 필요 * 100 FPM intake * 모든 duct와 plenum이 negative pressure임 * contaminated air가 흐르는 cabinet 내부의 공기통로의 Positive pressure는 negative pressure으로 둘러싸여 있어 내부공기가 누출되지 않음 * Chamber 내부에 vapour나 gas 발생시에도 100% 배출됨 Class III * 완전 밀폐형 Glove box * Chamber air 는 100% exhaust * Cabinet내부 negative pressure 유지 * 참고 : A1 or B1 타입은 현재 거의 추천하지 않습니다. NSF NSF International (The National Sanitation Foundation)에서 지정하는 cabinet 규정에 의하면 cabinet 위치 변경, HEPA filter 교환/수리 등이 이루어진 후에는 반드시 안정성 Test를 해야하고, 그렇지 않은 경우도 매년 안전성 Test를 해야한다. 242
Class ll Biological Safety Cabinets Steril GARD e3 < Class ll type A2 > NSF International Standard #49의 범위를 뛰어넘는 성능 유지 모든 제품의 출고시, 메이커에서 수행한 품질 검사 후 보고서 제공 기존의 SterilGARD III advanced제품에 비해 한층 더 향상된 기능. Bio Safety Level 3에 해당하는 미생물 취급 One-piece interior wall construction 특징 StediFLOW TM ; High-efficiency airflow control system : - 성능이나 안전도의 저하 없이 에너지 소비량을 줄임. - Blow motor의 과부하에 의한 열의 발생양을 감소. - 작동 시 소음 및 진동이 현저하게 감소. - 연속적이며 안전한 운전을 위해 motor의 자가 조절. - 7~10년 이상으로 HEPA filter의 수명을 연장해서 사용가능. StediFLOW TM VFD Motor controller - HEPA filter가 막혀서 발생하는 압력을, 통과하는 공기량의 손실 없이 약 250% 까지 자동으로 보정. - 공급되는 전원의 불안정으로 발생하는 Motor의 회전속도 변동 현상 차단. - 순환하는 공기량의 유지로 인해 성능의 감소를 줄이고 작업자와 시료의 보호하는 밸런스를 항상 유지 할 수 있다. - 기존 모델에 비해 소비되는 에너지를 약 40~60% 이상 감소. ReadySAFE TM mode : - 연속 작동 중 대기시 Viewscreen을 닫아도 bypass되는 공기 유입구 (Intake) 설계. - Viewscreen을 닫았을 때 Motor의 회전속도 30% 감소시켜 에너지 절약. - 지속적인 청정도 유지가 가능하여, 별도의 워밍업이 필요치 않음. - Viewscreen이 닫혔을 때 증가하는 motor의 부하 및 소음이 현저하게 감소. 243
Class ll Biological Safety Cabinets 혁신적인 UniPressure TM Preflow Plenum 디자인 : - HEPA filter에 걸리는 압력을 균등하게 배분함으로써 에너지 효율 및 필터의 수명을 증대시킬 수 있음. - 공급/배출 HEPA filter를 감싸고 있는 Plenum은 항상 음압에 둘러 싸여 있어서 오염원의 외부 노출을 막을 수 있음. - Plenum 및 Blow motor의 장착 디자인은 HEPA filter의 전면적에 압력을 고루 형성 시켜 작업공간에 항상 일정한 공기흐름을 유지 할 수 있게 함. - 기기의 전면부에서 접근 가능하며, Plenum과 Filter의 간격 조절이 가능하여 교체가 쉽다. Momentum Air Curtain : - 작업공간 보다 약 2배 이상 높은 유속의 Air curtain이 작업대 전면부에 형성. - 외부의 오염원이 내부로 침투하는 것을 막을 수 있음. High-Velocity Return Air slots 디자인 : - Viewscreen이 열리는 양 옆의 벽과 상단 프레임에 일렬로 Air slots 형성 - 고속의 흡입력을 유지하는 Air slots으로 돌발적인 외부 오염원의 유입을 완벽 차단 - 작업공간 내부의 오염원의 외부 유출 완벽 차단 Negative-pressure Double wall Plenums 디자인 : - 음압을 형성하는 공기통로는 작업공간의 전면부를 제외한 양 옆과 뒤쪽 까지 형성. - Baker에서만 유일하게 적용하는 디자인. - 기기의 노후화나 부식으로 인해 발생할 수 있는 오염원의 누출을 물리적으로 막음. 244
Class ll Biological Safety Cabinets Balanced Airflow and Exhaust ensure uniformity : - Downflow와 Inflow의 균형을 완벽하게 유지.작업대의 앞쪽과 뒤쪽 유입구의 형성은 공기의 흐름을 균일하게 유지 시켜 줌. Blow motor를 통해 양압이 형성된 공기는 70%가 재순환되고 30%가 Exhaust filter를 통해 배출됨. Slanted Viewscreen의 적용 : - 약 10도 기울어진 Viewscreen의 적용. - 눈부심을 줄이고 작업공간의 접근성 및 시인성이 확보되어 작업자의 편리성과 안전성을 보장. 효율적인 내부 조명의 적용으로 에너지 효율을 높임 Microprocessor control panel : - 작업자의 눈높이에 Membrane key pad형 Control panel적용. - 각각의 버튼은 아이콘형으로 구분 표시되어 조작이 쉽다. Exclusive Designs Simplify Certification and testing : - 장비의 유지 보수가 쉬운 내부 디자인의 적용으로 필터의 교체 및 공기 누출방지를 위한 특수 Gasket으로 완벽 밀폐 - Internal Damper를 통해 Exhaust와 Downflow사이의 밸런스를 쉽게 조절. - 큰 Magnehelic gauge를 통해 공기 유입구 및 필터의 막힘등 장비의 운전 상태를 확인. Performance envelope 적용 - 수차례의 Microbiological aerosol test에 의거하여 Intake와 Downflow의 밸런스가 유지되는 최적의 포인트를 설정한 후 작업자와 시료를 보호 할 수 있는 유효 범위를 산출. 옵션 외부 공기 배출 덕트와의 연결 (Option) FlexAIR Exhaust connection : - NSF49에서 권장하는 Class II Type A2 장비의 공기 배출 시스템이며, 건물 내 공기 조화 시스템의 문제로(예를 들면 건물 내 환기 팬의 고장 및 과부하로 인해 장비의 배출 속도나 배출양의 변동) 발생하는 장비의 밸런스가 무너지는 것을 외부 공기의 유입 보정으로 막을 수 있음. 필요 시 Air tight damper를 장착 할 수 있음. 245
Class ll Biological Safety Cabinets SterilchemGARD III Advance < Class II type B2 > - NSF International Standard #49의 범위를 뛰어넘는 성능 유지. - 모든 제품의 출고시, 메이커에서 수행한 품질 검사 후 보고서 제공. - Bio Safety Level 3와 4의 일부에 해당하는 미생물 취급. - 건물 내 덕트와 Hard connection 후 사용. - One-piece interior wall construction. 특징 Total Exhaust, No Recirculation: - Intake는 Prefilter와 Blow motor를 통해 Supply HEPA filter로 유입 되어 작업공간에 공급 된다. - 작업 공간을 거친 오염된 공기는 100% 배출 되며, 이때는 건물 내 배기 시스템에 연결되어 Exhaust air flow가 유지된다. - 위와 같은 개념으로 인해 소량의 gas나 vapour를 생성하는 시료의 작업도 가능하다. StediVOLT TM Motor Speed Controller : - 공급 되는 전압을 항상 일정하게 유지 시킴으로써 Blow motor의 회전속 도 또한 유지 시켜 주어 최적의 성능을 발휘. - Blow motor의 회전 속도 유지는 HEPA filter의 수명을 늘리는데 기여 한다. Blower/Motor system Model SG403A TX, 4 SG603A TX, 6 Motor Loading Capacity Pressure Drop Inceraes 200% 100% Maximum Increase With Manual Blower Speed Control 300% 180% - 7~10년 이상으로 HEPA filter의 수명을 연장해서 사용 가능. - HEPA filter의 막힘으로 인해 통과하는 공기량이감소하면 모터의 회전 속도를 모델별로 100%와 - 200%까지 자동 증가시켜 유지 하고, 수동으로 임의 조작까지 하면 180%와 300%까지 증가시켜 유지 할 수 있다. 246
Class ll Biological Safety Cabinets - 연속 작동 중 대기시 Viewscreen을 닫아도 공기 유입이 bypass되는 Padded Armrest를 적용 하였다. - 지속적인 청정도 유지가 가능하여, 별도의 워밍업이 필요치 않음. - Viewscreen이 닫혔을 때 증가하는 motor의 부하 및 소음이 현저하게 감소 혁신적인 UniPressureTM Preflow Plenum 디자인 : - HEPA filter에 걸리는 압력을 균등하게 배분함으로써 에너지 효율 및 필터의 수명을 증대시킬 수 있음. 공급/배출 HEPA filter를 감싸고 있는 Plenum은 항상 음압에 둘러 싸여 있어서 오염원의 외부 노출을 막을 수 있음. - Plenum 및 Blow motor의 장착 디자인은 HEPA filter의 전면적에 압력을 고루 형성 시켜 작업공간에 항상 일정한 공기흐름을 유지 할 수 있게 함. - 기기의 전면부에서 접근 가능하며, Plenum과 Filter의 간격 조절이 가능하여 교체가 쉽다. Momentum Air Curtain : - 작업공간 보다 약 2배 이상 높은 유속의 Aircurtain이 작업대 전면부에 형성. - 외부의 오염원이 내부로 침투하는 것을 막을 수 있음. High-Velocity Return Air slots 디자인 : - Viewscreen이 열리는 양 옆의 벽과 상단 프레임에 일렬로 Air slots 형성. - 고속의 흡입력을 유지하는 Air slots으로 돌발적인 외부 오염원의 유입을 완벽 차단. - 작업공간 내부의 오염원의 외부 유출 완벽 차단 Negative-pressure Double wall Plenums 디자인 : - 음압을 형성하는 공기통로는 작업공간의 전면부를 제외한 양 옆과 뒤쪽 까지 형성. - Baker에서만 유일하게 적용하는 디자인. - 기기의 노후화나 부식으로 인해 발생할 수 있는 오염원의 누출을 물리적으로 막음. 247
Class ll Biological Safety Cabinets Slanted Viewscreen의 적용 : - 약 10도 기울어진 Viewscreen의 적용. 눈부심을 줄이고 작업공간의 접근성 및 시인성이 확보되어 작업자의 편리성과 안전성을 보장. 효율적인 내부 조명의 적용으로 에너지 효율을 높임 Microprocessor control panel : - 작업자의 눈높이에 Membrane key pad형 Control panel적용 각각의 버튼은 아이콘형으로 구분 표시되어 조작이 쉽다. AirExclusive Designs Simplify Certification and testing : - 장비의 유지 보수가 쉬운 내부 디자인의 적용으로 필터의 교체 및 공기 누출 방지 Gasket 적용 Airflow Monitor, Automatic Shut-off : - Mass airflow 알람 시스템은 배출량이 설정보다 약 20% 이하로 감소되면 알람을 발생 시킨다. - 설정보다 낮은 공기 흐름이 6초 이상 지속되면 알람 시스템은 작업공간의 양압화를 막기위해 외부에서 공기를 공급하는 장비 내부의 Blow motor의 작동을 차단한다. - 건물 내 덕트로의 공기 배출이 원활하지 않아 알람이 발생 할 때, 장비의 작동은 멈추게 되며 사용이 중지된다. 옵션 외부 공기 배출 덕트와의 연결 Exhaust Transition (ET) for Hard connection : - SG403/603 ATX 모델에는 건물 내 환기 시스템과 연결 할 수 있는 연결부등이 같이 공급 된다. 장비의 Exhaust 부분에 Airflow monitor probe를 장착하는 사각형 판넬과 1번(기본 공급) 또는 2번(옵션 사항)의 덕트 연결부를 적용 할 수 있다. 248
Class ll Biological Safety Cabinets 참고 : BSC 타입별 외부 공기 배출 덕트와의 연결 (NSF49에 의거) SterilGARD III (Class II type A2) - 본 장비는 자체 Blow motor에 의해 70%가 재 순환 되고, 30%가 배출 된다. 건물 내 배기 덕트와 연결하여 사용 시, 건물 공조 시스템에 장착된 Fan의 문제로 인해 장비의 공기 배출구에서 역류 현상이 발생 하거나, 과도하게 많이 배출되는 현상이 발생할 수 있다. 이런 현상이 발생하면 장비 내부의 Downflow 와 Intake의 밸런스가 무너지면서 시료 및 작업자의 안전을 보장하는 범위를 벗어날 가능성이 있다. 따라서 Class II type A2 장비의 외부 공기 배출은 Canopy connection을 사용하도록 권장 하고 있다. SterilchemGARD III (Class II type B2) - 본 장비는 100% 외부 배출형 이다. 장비에 장착된 Blow motor는 작업공간에 공기를 공급 하는 역할만 수행 하고, 배출은 모두 건물 공조 시스템에 장착된 Fan에 의해 이뤄진다. 따라서 기본적으로 장비의 Blow motor에서 공급 하는 압력 보다, 건물 내 Fan에서 빨아 당기는 압력이 더 높게 유지 되어야 한다. 이러한 차이가 무너지면 장비에서 작업하는 시료가 실험실 환경으로 여과없이 누출되기 때문에 기기의 작동을 차단한다. 기본적으로 Hard connection을 구성해야 하며, Air tight damper를 통해 배출량을 조절 하는 것이 좋다. Specification (220V / 60Hz 기준) SterilGARD e3 (Class II Type A2) SterilchemGARD III (Class II Type B2) Models SG403A-HE-INT SG503A-HE-INT SG603A-HE-INT SG403A-TX-INT SG603A-TX-INT Size 4 ft. 5 ft. 6 ft. 4 ft. 6 ft. Electrical 220V / 60Hz Amps/Fuse (내부 콘센트 제외 ) 3/6.3 A 4/6.3 A 5/6.3 A 4/6.3 A 5/6.3 A BTU/hr (Sash 8 open) 1474 1638 2743 1571 3063 Interior Dimensions(mm) 1168w x 512.8f-b x 697h 1473.2w x 498.5f-b x 697h 1778w x 498.5f-b x 697h 1168w x 462f-b x 697.4h 1778w x 462f-b x 697.4h Exterior Dimensions(mm) (스탠드 포함) Net weight 1366w x 777f-b x 2283h 668 lbs 1670w x 777f-b x 2283h 756 lbs 1978w x 777f-b x 2283h 830 lbs 1364.4w x 776.3f-b x 2362h 596 lbs 1977.3w x 776.3f-b x 2362h 811 lbs Ship weight 798 lbs 906 lbs 1000 lbs 798 lbs 1063 lbs Lighting 32W x 2 ea 32W x 2 ea 32W x 2 ea 32W x 2 ea 32W x 2 ea UV lamp 40W x 1 ea 40W x 1 ea 75W x 1 ea 40W x 1 ea 75W x 1 ea 내부 콘센트 220V x 2 ea 220V x 2 ea 220V x 2 ea 220V x 2 ea 220V x 2 ea Protection level (박테리아) Biosafety level 3 (실험자/시료/환경 보호) Protection from gas & vapour No 실험자/시료보호가능 Protection from gas & vapour (건물 내 덕트와 연결시) Air flow monitor system 실험자/환경 보호 가능 Option 실험자/시료/환경 보호 가능 Standard Exhaust Requirements FlexAIR canopy exhaust connection(cec) BSC model Sash Height Exhaust Flow FlexAir Static Pressure Required Rate (CFM) (inches Water Column) Duct dia. : 8" Duct dia. : 10" Duct dia. : 12" SG-403A-HE 8" 290 (137L/sec) 0.18 (45 pa) 0.11 (27 pa) 0.08 (20 pa) SG-503A-HE 360 (170L/sec) 0.27 (68 pa) 0.16 (40 pa) 0.12 (30 pa) SG-603A-HE 460 (217L/sec) 0.31 (78 pa) 0.18 (45 pa) 0.13 (33 pa) Hard exhaust connection(hec) BSC model Sash Height Exhaust Flow Pressure Required Rate (CFM) (inches Water Column) Clean Exhaust Filte 50% Loaded Filter SG-403A-TX 8" 750 ~ 800 1.2 (299 pa) 1.6 (398 pa) (354~378L/sec) G-603A-TX 1150 ~ 1200 11.5 (374 pa) 2.0 (498 pa) (543~566L/sec) 249
Class III Biological Safety Cabinets IsoGARD < Class III BSC > - NSF International Standard #49의 범위를 뛰어넘는 성능 유지. - 모든 제품의 출고시, 메이커에서 수행한 품질 검사 후 보고서 제공. - Bio Safety Level 4에 해당하는 미생물까지 취급. 모든 BSC제품 중 최고 레벨의 안전성 보장. - One-piece interior wall construction. 특징 실험자 보호 Gas-tight 구조 및 Pass-thru chamber의 이용으로 고위험군 시료에 노출되지 않는다. 작동 시 작업공간의 내부는 0.5 Water column pressure를 유지 시료의 보호 작업 공간에 공급되는 공기는 모두 Full size HEPA filter를 통해 여과되어 공급 되며, 시료에서 발생하는 Particle을 모두 배출하도록 유도한다. (Unidirectional Air Flow) 실험실 환경의 보호 Baker만의 독보적인 Plug-and seal canister style의 HEPA exhaust filter 적용. 여과된 공기는 건물 배기 시스템을 통해 처리. 인체 공학적 디자인 작업의 효율 및 편리성을 제공하기 위해 10도 기울어진 View-screen 적용하였다. Exhaust plenum은 작업자의 무릎이하에 위치 하며, Cabinet 외부에 위치하는 조명은 작업자의 머리 위에 위치한다. IsoGARD 내부는 SUS-316L 타입의 재질이 적용 되었으며, 둥근 모서리를 적용하였고 나사나 볼트등을 사용하지 않아 유지 관리가 용이하다. 작업공간의 조명은 외부 canopy에 장착이 되며, 내부 재질은 눈부심이 적게 표면 처리 되어 있다. Digital pressure gauge 및 시각/청각 알람 시스템이 장착되어 작동 압력 등을 모니터링 하며, 작동 상황에 대한 인지가 빠르다. View screen은 Safety glass로 이뤄져 있고, Solid closed-cell silicone gasket과 neoprene glove가 적용되었다. Exhaust Requirement 건물 내 HEPA Filter 배기 시스템 건물 배기 시스템은 적어도 음압으로 흡기되어야 하며, 약 2.5 water column static pressure를 유지해야 한다. 250
Class III Biological Safety Cabinets 옵션 Top hinged viewscreen 작업 공간에 큰 장비등을 넣을 수 있도록 전면부가 위로 열리는 타입 Disinfectant dunk tank 오염 되지 않은 표면을 유지하며, 장비의 내부에서 바로 폐기물 처리 오염된 폐기물을 외부 유출 없이 바로 멸균 Specification (115V / 60Hz 기준) Models Size Interior Dimensions(mm) Exterior Dimensions(mm) Net weight hip weight Exhaust Requirement IsoGARD (Class III Glove box type) IG-14 4-Glove Pass-thru : 1016w x 452.1f-b Working area : 1852w x 452.1f-b 2590w x 841.4f-b x 2159h (스탠드 포함) 950 lbs 1150 lbs 350 CFM at negative 2.5 ± water column static pressure 251
Pharmacy Barrier Isolators ChemoSHIELD[negative pressure]glovebox Glovebox work area 내부가 negative pressure로 되어 있어 hazardous or potent pharmaceutical compound, Chemotherapy Agents, IV admixture등의 안전한 조작을 위해 제작됨. 특징 Unindirectional HEPA-filter airflow - 단일방향으로 air flow되며 이것은 piston effect 로 work area 및 pass-thru chamber를 class 100 이상의 공기청정상태로 유지함. Negative Pressure - Glovebox의 Main chamber와 투입구 (pass-thru chamber)는 모두 negative pressure로 되어 있으며, pass-thru chamber가 main chamber에 비해 약간 더 강한 negative pressure로 되어 있어 이는 Main chamber 내부 sample의 안정성을 더 고려한 사항임. (Standard version) < ChemoSHIELD Glovebox > Ergonomic 디자인 - 36 (seated), 45 (standing)의 2가지 높이로 work space를 조절 가능하고, Front viewscreen이 10도 기울어져 있어 사용자의 편리성을 고려했으며, 상단에는 경첩이 달려있어 실험장비 등을 투입구를 통하지 않고 직접 작업대로 넣을 수 있다. 또한 main chamber와 투입구에 glove가 있어 내부 곳곳에 손을 넣어 작업하기 편하고, 내부 cabinet의 모서리는 7/16 반경의 곡면으로 처리되어 청소하기 편리함. Pass-thru chamber - 외부 interchange door는 내부의 interchange door가 닫혀야만 열리는 구조로 되어 있어 안전하고, 내부 interchange door는 수직으로 여닫는 구조로 되어 있어 sample의 이동이 용이함. 46 W x 24 D 의넓은작업대 Main chamber와 interchange chamber의 내부는 Type 316 stainless steel로 되어 있어 부식에 대한 저항성이 강함. Glovebox의 외장은 긁힘에 저항성이 강한 PermaWhite baked enamel 로 마감처리되어 강한 내구성 지님. 형광등이 Ballast와 함께 cabinet 외부에 위치하여 chamber 내부 온도 상승을 억제함. Main chamber 내부에 IV bar가 있어 IV bag 또는 병을 걸어두는데 편리함. Cabinet 내부에 Blower가 없어 소음이 없음. Exhaust Requirement - 8 Flange Hard Connection, Minimum 375CFM and -2.0 WC static pressure의 Exhaust system이 필요함. Specification Cat. No. CS600 Size 6 ft. Amps Ballast 1.0 Total 1.0 Exhaust Connection 375CFM with static pressure- -2.0 wc at exhaust Cat. No. CS600 Interior Dimensions Interchange (inch) 171/2 W x 24 F-B x 25 H Main Chamber (inch) 46 W x 24 F-B x 25 H Exterior Dimensions (inch) 75 W x 35 F-B x 90 H Net Weight 615 lbs Ship Weight 815 lbs 252
Laboratory Animal Research Product AniGARD ll Animal Transfer Station Class 100, Vertical Flow Recirculating Clean Bench < AniGARD II - AG-4-ATS > 실험실 내에서 실험동물의 cage를 교환해 줄 때 외부의 노출을 최소화하는 상 태에서 진행할 수 있도록 고안된 cabinet. 3면에서 작업을 할 수 있기에 실용적이며, 인체공학적으로 디자인 되어 있어 사용자의 편의를 생각한 cabinet. 특징 Improved Productivity - 가볍고 compact한 디자인과 standard caster 및 optional pull bar는 실험실 내에서 cabinet의 운반을 편리하게 해주고, 평평한 작업공간은 cage의 이동에 용이하며, 3면의 작업공간으로 팀웍으로 하는 멀티 작업이 가능함. Access Opening On 3 Sides 70 FPM 175 FPM Diffuser Exclusive Momentum Air Curtain - Stainless steel diffuser가 supply filter의 바로아래에 위치하여 viewscreen의 안쪽면에 175FPM의 빠른 airflow가 momentum air curtain을 형성하여 캐비닛 외부에서 내부로 오염된 공기가 침투할 수 없는 방벽을 생성시 키고, work surface 둘레 가장자리의 air slot을 통해 contaminated air가 intake 됨. Airflow Design Assures Maximum Efficiency - Work area air와 room air가 work surface air slot으로 intake 된 후 작업대 밑에 위치한 pre-filter와 primary HEPA filter를 거쳐 1차 filtering 된 후 cabinet 상 단의 secondary HEPA filter를 거쳐 작업대 내로 들어오는 이중 HEPA filter 방식을 도입함으로써 cross-contamination과 productcontamination을 막아줌. Double HEPA filter로 ISO Class 100이상의 뛰어난 공기 청정도를 유지. Ergonomic Benefits - 10도 기울어진 viewscreen으로 사용자의 편리성 도모, 적당한 밝기로 눈의 피로 감소, 3면에서 작업이 가능한 실용성, 36"에 서 48"까지 작업대 높이 조절 가능, 저소음 실현, 경량으로 운반 용이. EPA-Filtered Air -Negative Pressure EPA-Filtered Air +Positive Pressure Contaminated Air From Work Surface Room Air 작업대 아래의 pre-filter는 쉽게 분리해서 세척이 가능하므로 HEPA filter 의 수명 연장에 기여. Blower 및 모든 electrical parts가 clean air plenum 내부의 primary HEPA filter의 아래에 위치하여 오염에 노출될 확률을 줄임으로써 부품의 수명을 길게 해주며, maintenance 비용을 절약해 줌. Specification Cat. No. AG-4-ATS Size 4 ft. Access opening(working position) 12 Access opening(loading/cleaning position) 21 Electrical 230V,AC Amps/Breaker 10.3 / 15 Minimum exterior dimensions(h x F-B x W) 723/6 x 30 x 52 Interior Dimensions 221/2 x 24 x 443/4 Net weight 425 lbs Ship weight 550 lbs Warranty 36 months 253
Laboratory Animal Research Product AniGARD VF Animal Transfer Station 실험실 내에서 실험동물의 cage를 교환해 주는 동안 animal protection과 particulate control을 제공해주는 class 100, vertical flow recirculating type의 clean bench 특징 Improved Productivity and Optimal Mobile Operation - Optional caster와 optional pull bar는 실험실 내에서 cabinet의 운반을 편리하게 해주고, 평평한 작업공간은 cage의 이동이 용이하며, 21"~40"까지 높이 조절이 가능해져 낮은 높이로 좁은 문이나 좁은 공간에 두기 편리함. Exclusive Momentum Air Curtain and High Velocity Air Slots - Stainless steel diffuser가 supply filter의 바로 아래에 위치하여 viewscreen의 안쪽으로 175FPM의 airflow를 형성하고, 안쪽으로는 50~60FPM을 형성하여 작업대 주변에 외부에서 꿰뚫을 수 없는 방벽을 생성시키며, 작업판 주변의 구멍을 통해 외부의 공기가 빨려 들어감. < AniGARD VF - AG-5-VF > Room Intake Air to HEPA Filter Re-circulated Air HEPA-Filtered Downflow Air Laboratory Air Captured by High Velocity Return Air Slots Discharged Air Ergonomic Benefits - 10도 기울어진 viewscreen으로 사용자의 편리성 도모, 적당한 밝기로 눈의 피로 감소, 작업대의 높이를 21"~40"으로 조절 가능하며, 저소음, 경량으로 운반 용이. Primary HEPA filter는 front plenum area에서 부터 쉽게 분리됨. Hinged viewscreen은 필요할 때 열어놓을 수 있는 특이한 latch system 기능이 있음. Specification Cat. No. AG-5-VF Size 5 ft. Access opening (working position) 14 Access opening (loading/cleaning position) 301/2 Electrical 230V, AC Amps/Breaker 12/15, 14/20 with lift Exterior dimension range (stationery stand) Work surpace height (stationery stand) Interior Dimensions Ship weight Warranty 887/8 ~ 981/8 H x 32 F-B x 651/2 W 301/4 to 383/4 W 58 H x 23 F-B x 313/4 W 800 lbs 36 months 254