ISSN Review Article Review of recent developments for intra-oral scanners Jong-Hoon Choi 1, Y

Size: px
Start display at page:

Download "ISSN Review Article Review of recent developments for intra-oral scanners Jong-Hoon Choi 1, Y"

Transcription

1 ISSN Review Article Review of recent developments for intra-oral scanners Jong-Hoon Choi 1, Young-Jun Lim 2, Won-Jin Lee 3, Jung-Suk Han 2, Seung-Pyo Lee 4 * 1 Department of Dentistry, School of Dentistry, Seoul National University, Seoul, Republic of Korea 2 Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea 3 Department of Oral Maxillofacial Radiology, School of Dentistry, Seoul National University, Seoul, Republic of Korea 4 Department of Oral Anatomy, School of Dentistry, Seoul National University, Seoul, Republic of Korea Making a model that is an accurate replica of the oral structure requires precision and efficiency. Nowadays, rapid technological advances bring digitalization in dentistry. One of the most important works in digital dentistry is three-dimensional modeling of the oral cavity and digitizing the 3D data. Among the three components of CAD/CAM, 1 data capture component (digitizers), 2 design component (CAD software), 3 manufacturing component (CAM), the basic component that has a significant impact on the other processes is the data capture component, i.e. intra-oral scanners. This literature review discusses the principles and clinical use of intra-oral scanners in dentistry based on recent publications of the past 5 years using the PubMed and Google Scholar databases. (J Dent Rehabil Appl Sci 2015;31(2):112-25) Key words: intra-oral scanner; digital dental impression; three-dimensional imaging 서론 구강내의모습을재현하는복제모델을만드는것은치과진료에서가장중요한과정중하나이며, 정확성과효율성이동시에만족되어야하는과정이다. 현재대부분의치과에서는구강내모습을재현할때전통적인인상채득방법을사용하고있다. 하지만디지털기술이적용되면서치과진료에서도기존의방법을대체하는디지털화된임상방법이개발되어사용되고있다. 이제치아삭제를제외한모든과정, 즉인상, 모델제작, 보철물제작에이르기까지모두컴퓨터와기계가담당할수있게되었다. 이러한것을가능하게하는가장중요한작 업중하나가바로구강내의모습을 3차원적으로재구성하여디지털화하는것이다. CAD/CAM 시스템의 3 가지성분 1 data capture component (digitizers), 2 design component (CAD software), 3 manufacturing component (CAM) 중에서가장기본이되고뒤의과정에상당한영향을끼치는성분은 data capture component, 즉구강스캐너이다. 본종설은현재시판중인 4가지종류 (CEREC, itero, LAVA C.O.S., Trios ) 의구내스캐너각각의구동원리와정확성에관하여알아보았고, 현재구강스캐너가치과영역에서임상적으로적용되고있는분야와그분야에서의적용정도를분석하는데목적이있다. *Correspondence to: Seung-Pyo Lee Professor, Department of Oral Anatomy, School of Dentistry, Seoul National University, 28 Yeongeon-dong, Jongro-gu, Seoul, , Republic of Korea Tel: , Fax: , orana9@snu.ac.kr Received: February 12, 2015/Last Revision: May 15, 2015/Accepted: May 31, 2015 Copyright 2015 The Korean Academy of Stomatognathic Function and Occlusion. cc It is identical to Creative Commons Non-Commercial License. 112

2 Review of recent developments for intra-oral scanners 1. 구강스캐너의작동원리와각스캐너의특성 (1) 삼각법 (triangulation technique) 삼각법이란물체의형태에대한 3 차원데이터를모아 3 차원모델을만드는데사용되는비접촉성기술중하 나이다. 1 수동적삼각법과능동적삼각법이있고모두 사용된다. 먼저능동적삼각법은빛을직접방출하여물 체의표면에투영하고그반사된빛을다시수집하여목 표물체의위치를계산해낸다. 반면수동적삼각법에서 는스캐너에서직접빛을방출하지않고반사된주위의 빛을탐지해낸다. 수동적 (passive) 삼각법은다른말로수동적스테레오 비전이라고부르며사진측량알고리즘을사용한다. 2 이 기술은위치와각도를이미알고있는두개의카메라가 두개의입체이미지를얻는것을기초로한다. 3 차원의 한점을두카메라로부터얻은영상정보에서카메라의 위치와이를보는영상점의두벡터는공통평면상에존 재한다는것이등극선기하인데, 등극선의일치하는점 을통해앞서얻은두개의입체이미지에일치하는물 체의위치를알아내는것이다. 수동적삼각법은매우높은정확성을제공하지만물 체가높은대조도를가지고있어야하고모서리가뚜렷하게형성되어있어야한다. 카메라를 3개사용하면더욱더정확해질수있다. 3 목표물이아닌것들은전혀측정되지않으며조절된빛을사용하지않을경우에는주위의빛이물체의상을올바르게얻어내는능력에매우영향을준다. 이방법의장점은구성성분이몇개없고매우싸며작동원리또한우리의눈과같아서간단하다는것이다. 4 Fig. 1과같이, 능동적 (active) 삼각법은레이저에의해생성된빛광선이거울에반사되어원하는물체 (Object) 로비추게된다. 이기술은카메라 (Position Detector, Lens) 에서나온실선과물체가만나는지점인레이저도트와카메라그리고레이저가삼각형을이루게된다. 카메라와레이저사이의거리를알고있고이것을기준선거리 (d) 라고부르며삼각형의한변에해당한다. 레이저의각도또한알려져있다 (θ). 5 카메라시야에서얻을수있는카메라각도 (Φ) 를측정해낸후기준선거리와레이저각도와함께 3가지정보를가지고삼각형의모양과크기를결정할수있으며삼각형의한점인레이저도트의 3차원적위치 (X, Y, Z) 를삼각계산을통해알아낼수있다. 대부분의경우에서는레이저도트한점만을사용하지않고레이저광선이나약속된레이저패턴을사용하여물체에대한정보를더빠르게얻는다. Laser Object Deflector θ Z Position Detector d Mirror (X,Z) P p Lens Φ x f 0 X Fig. 1. Active triangulation method. 7 J Dent Rehabil Appl Sci 2015;31(2):

3 Choi JH, Lim YJ, Lee WJ, Han JS, Lee SP 모든삼각법에가장큰문제는언제나목표로하는표면이레이저와카메라에측정되지않을가능성이있다는점이다. 이론적인해결법은삼각법의각도를최대한줄이는것이지만삼각법의각도를줄이게되면그만큼정확성이떨어지게된다. 6 따라서정확성과측정되지않을가능성에대한바람직한균형이고려되어야만한다. 광삼각 3차원이미지측정장치는레이저빔이나오는부분, 카메라영상센서부분, 이송장치, 영상처리컴퓨터부로이루어져있다. 7 레이저빔이이송장치위에있는목표물에조사되고이송장치를통해물체를이동시키면서카메라영상센서로영상을획득하여컴퓨터로보내게되면컴퓨터가정보를수집하여 3차원이미지를만들어내어화면에나타나게된다. CEREC by Sirona dental system (Bensheim, Germany) CEREC은 Chairside Economical Restoration of Esthetic Ceramics 의줄임말로 3차원디지털스캐너와밀링머신을합쳐놓아한번의내원으로도재블록을가공하여체어사이드에서치과적인수복을해줄수있도록만들어진기계이다. 한번의내원으로바로수복을할수있게되면여러번내원하지않아도된다. 그러므로임시치아를만들필요도없고그에파생되는여러문제점들도겪지않아도된다. CEREC 시스템은 3차원디지털화를하기위하여컴퓨터를이용한기술을사용하게되는데디지털모델로서정보를저장하고인접치와 대합치와의생체적인비교를통해모양을정하고치과의사로하여금모양을디자인할수있게해준다. 디자인을정하고나면밀링머신으로정보를전송하여도재나합성재료블록을가지고단일크라운을약 4분만에만들수있도록설계되었다. 요즘은인레이, 온레이, 크라운은물론라미네이트비니어, 심지어브릿지까지한번의내원으로완성할수있다. 8 CEREC 시스템은앞에설명했던능동적삼각법의원리로이미지를얻는다. 9 먼저프랩된치아주변의조직을적당히당기거나정리한후지혈을하여마진이잘보이게해야한다. 그리고청색광패턴이무작위적으로변화하는기구를이용하여빛을물체에조사한다. 그다음, 각각의얻은데이터를가지고광학적빛의특성과이미지평면사이의거리사이의관계를얻어내어이미지를만들게된다. 10 Fig. 2를보면빛근원은 7.1, 7.2, 7.3에해당하는조명광선을만들어내며이것은목표물체인 2을향하여나아가게된다. 6의이미지센서가목표물체 2에반사된반사광선 (9.1, 9.2, 9.3) 을받게되는데이때 focusing system(5) 을거치게되면반사광선이이미지센서 (6) 로잘모이게된다. 사실빛근원 (3) 은빛의강도면에서독립적으로조절되는다양한구성요소들의조합 (3.1, 3.2, 3.3) 으로이루어져있다. 따라서각각의센서요소가감지하는빛의강도는스캐너의머리부분과목표사물에해당하는점사이의직접적인거리를표시해준다. 11 이때이 CEREC 시스템에서중요한부분인광삼각법을사용하기위해서는반사하는표면이일정해야한다. 일 S Fig. 2. CEREC scanning principle J Dent Rehabil Appl Sci 2015;31(2):112-25

4 Review of recent developments for intra-oral scanners 정하지않다면법랑질, 상아질, 아말감, 레진, 치주조직등이모두다르게빛을반사시키게될것이다. 따라서스캔작업을시행하기전에생체적합적인 TITANIUM DIOXIDE 파우더를완전하게덮어서표면의반사능을동일하게만들어야한다. 9 (2) 공초점레이저현미경 (confocal laser scanner microscopy) 공초점레이저현미경 (confocal laser scanner microscopy, CLSM) 은선택된깊이에서의집중된영상을얻을수있는기술로 optical sectioning이라고알려진과정을거친다. Optical sectioning이란깊이를선택하고그곳에초점이맞지않는신호는광검출기앞에있는공초점조리개로제거하여초점면에맞는신호만을모으는작업이다. 12 이미지들을각각의포인트들로부터모아컴퓨터로 3차원이미지를만들어낸다. 이기술을사용하게되면불투명한표본의표면모습과투명한표본의내부이미지를얻을수있게된다. 전통적인현미경이빛이통과하는만큼표본을관찰했다면공초점현미경은단지한번에한깊이의이미지들을볼수있으며그깊이와초점을조절할수있다. 공초점현미경의원리는 Marvin Minsky가 1961년에특허를냈지만레이저를발전시키고지금의형태를갖는 CLSM을만드는데까지는 30년정도가걸렸다. CLSM 기술에서레이저광선은조리개를지나게되고이것은대물렌즈를지나면서작은초점의용적을갖도록모이게되며, 우리가원하는물체의내부나표면에위치하게된다. 흩뿌려지고반사된레이저빛은다시대물렌즈를지나게되고그빛중일부분은빔분리기에의해탐지장치로보내지게된다. 이장치는형광성공초점현미경에서사용될때는필터를가지고있는데이필터는선택적으로형광성파장의빛은통과시키고원래방출된파장의빛은차단시킨다. 핀홀을지나고난후빛의세기는영상탐지기구에의해탐지되어서빛신호를전기신호로바꾸어컴퓨터가이를녹화하게한다. 제한된검출조리개는초점에서오지않은빛은다차단해버리고, 돌아오는빛의대부분은핀홀에의해차단되어서초점에맞지않는빛은다억제된다. 그결과로전통적인형광현미경기술에서얻은이미지보다더세밀한이미지를얻을수있으며샘플에서다양한깊이에위치하는면들의이미지를얻을수있게된다. 13 조명을받은물체에서나온빛을탐지하여얻은결과이미지는하나의픽셀을이루게된다. 이미지픽셀은탐지된빛의상대적강도와일치하게된다. 하나나그이상의보조로조절되는진동거울을사용하여수평면에서빛의광선은샘플을따라스캔하게된다. 12 스캔을느리게할수록노이즈보다신호가더많이잡히며, 더나은대조도와더높은해상도를가지게된다. 현미경단이나대물렌즈를올리고내려서서로다른초점면으로부터의정보를얻을수있게된다. 컴퓨터는이연속적인초점면으로부터얻은 2차원정보를결합하여 3차원이미지로생성하게된다 itero by CADENT LTD (Tel Aviv, Israel) Cadent itero 디지털인상시스템은 2007년초에시장에나왔다. itero 시스템은앞서설명한평행하는공초점영상기술을사용한다. 13 Fig. 3을보면적색레이저광선세트 (38) 를초점광학 (42) 에통과시켜치아에조사한다. 이초점광학은모터 (72) 에의해위치가조절되며하나나그이상의초점면을정해준다. 치아에조사된광선은구조물에점을만들고, 반사된빛의강도는초점면의다양한위치에서측정된다. 치아의 3차원이미지는반사된빛의최대강도를통해알아낸점의위치를기초로하여만들어진다. 14 이기술로 itero는구강내모든구조물의이미지를만들어낼수있으며이때치아에는어떤코팅도하지않아도된다. 15 전체구조의 3차원이미지는그치아주변에서얻은표면의형상을모아만들게된다. 16 파우더를코팅하지않아도된다는큰장점이있다. 하지만색이있는이미지를얻기위해서추가적으로스캐너의머리부분에색상환을달아야하므로다른스캐너들보다스캐너의머리부분이크다는단점이있다. 8 이시스템에서는 3차원치아구조의 2차원색이미지역시, 구조를알려고할때와마찬가지로같은각도와방향을사용한다. 그결과각각의 2차원색이미지의 X-Y 좌표는 3차원스캔의 X-Y좌표와일치하는점을갖게된다. 색이미지는적색, 녹색, 청색의보색을사용하여얻어지며각각의단색들이합쳐져서여러가지색을갖는이미지를만들게된다. 17 세가지색광선은모두같은백색광에서얻어지며이색필터는회전이가능한디스크에한부분을차지하며모터와연결되어있다. 실제인상을뜨기위해서는프랩된치아를다섯번스캔한다. 교합면, 설측, 협측그리고인접치아의치간접촉점을차례로스캔하게되며약 15-20초정도시간이소요된 J Dent Rehabil Appl Sci 2015;31(2):

5 Choi JH, Lim YJ, Lee WJ, Han JS, Lee SP 31 semiconductor Laser polarizer Optic expander grating or microlens array polarizer imaging Optics telecentric main confocal optics motor relay Optics Z X Y endoscope modulation 70 ccd control A B Fig. 3. itero system. 16 다. 8 그리고남은치아를 45도각도로협측과설측으로스캔하여 4분의 1악이나대합치반악을스캔하게된다. 이러한스캔과정이끝난후환자의치아를중심교합으로위치시킨후스캔한다. 이모든과정을끝내는데약 3분정도가소요된다 TRIOS by 3shape A/S (Copenhagen, Denmark) 2010년 12월, 3shape 사에서 TRIOS를시장에출시했다. TRIOS는 itero와마찬가지로공초점현미경의원리를사용하며빠른스캔속도를가지고있다. 19 빛진동을물체에일으킬수있게하는조명방식을가지고있다. 그방식속에서진동은공간적이며다양한시간대를가지고있다. 이시스템은물체와스캐너사이의공간적인관계는그대로두면서초점면위치의범위에따라그방식의초점면의다양성을생성할수있는능력이있다. 다시말해스캐너를움직이지않고도진동을일으키는방식덕분에다양한초점면을생성할수있다는것이다. 또다양한시간대를가지는진동이적용될때한번의스캔으로초점면의서로다른위치들과서로다른거리에있는 2차원이미지를얻을수있다. 하나의초점면은스 캔된표면에어떤픽셀의위치와일치할때그패턴은표면의한점에초점이맞게조사되고높은대조도를가지기때문에픽셀값의큰진폭의상승을보이게된다. 19 각각의픽셀이초점이맞을수있는초점면을맞출수있는특정한설정을확인하는것이가능하다. 그러므로초점면의위치에따른대조도정보를 3차원표면정보로바꾸는것이가능하다. 스캔된물체의 3차원은카메라의센서세트 (180, 181) 에속한각각의센서에의해측정된정점과일치하는면을찾는것으로결정된다 (Fig. 4). 20 TRIOS의가장중요한특징은물체를스캔할때스캐너를움직이지않고다양한초점면을얻을수있다는점이다. 초점면은미리설정된진동수를가진주기적인방식으로연속적으로얻어져야한다. 그리고패턴을만들어내는수단 (130), 카메라 (180), 광학시스템 (150) 은서로긴밀하게연결되어스캔작업을해야한다. 20 더나아가 3차원표면을얻어내는시간은 probe와치아사이의상대적인움직임의영향을줄이기위해서매우작아져야한다. TRIOS는스캔되는물체의공간에서볼때 telecentricity의특성을가지고있고이러한특성과확대능이유지되면서초점면을변화시킬수있다. 116 J Dent Rehabil Appl Sci 2015;31(2):112-25

6 Review of recent developments for intra-oral scanners Fig. 4. Trios scanning system. 22 Object Camera Center Rotating off-axis Aperture Lens Rotating Image Point Image Plane Fig. 5. Active wavefront sampling principle. 23 (3) 활성파면샘플링 (active wavefront sampling and devices) 활성파면샘플링 (AWS) 은 3차원이미지를만드는기술로하나의카메라와활성파면샘플링의모듈만을사용한다. Fig. 5처럼, 활성파면샘플링모듈의가장단순한모습은측면조리개이며광축을주위로원을이루며움직이게된다. 이러한움직임은이미지평면에있는원에목표지점의회전을생산한다. 목표지점의깊이에대한정보는각각의포인트에서만들어지고원을이루고 있는포인트패턴들의반지름으로부터얻을수있게되며, 이러한것은회전하는활성파면샘플링모듈에의해생성되고있다. 따라서활성파면샘플링이미지는디지털카메라하나를가지고어떤시스템이든지다 3 차원으로작용할수있게할능력이있다. 21 그러므로여러개의카메라로 3차원기하를얻으려고노력할필요가없어진다. 명백하게활성파면샘플링시스템의가장중요한특징은바로깊이정보를얻기위해서단하나의광학로만있으면된다는것이다. J Dent Rehabil Appl Sci 2015;31(2):

7 Choi JH, Lim YJ, Lee WJ, Han JS, Lee SP Lava Chairside Oral Scanner by 3M ESPE (St Paul, MN, USA) Lava Chairside Oral Scanner (C.O.S.) 는 2006년 10월 3M ESPE 사에서처음으로만들어졌으며공식적으로출시된것은 2008년 2월이다. Lava C.O.S. 시스템은하나의모바일카트에 CPU, 터치스크린디스플레이, 스캐닝기구로구성되어있다. 22 Lava C.O.S. 카메라는매우복잡한광학시스템으로이루어져있는데 22개의렌즈시스템과 192개의파란 LED cell로구성되어있다. 그리고구강스캐너는 13.2 mm의폭과 390 g의무게가나간다. 23 Lava C.O.S. 는 3차원데이터를앞서말한활성파면샘플링의원리를이용하여대상에빛을조사하여얻어낸다고알려져있다. 이러한스캐닝방법을 3M ESPE사에서는 3D-in Motion technology 라고명명하고있다. 22 이스캐닝시스템은활성화된 3차원이미지시스템을제공하는데, 이것은회전하는측면조리개를포함하고있다. 이회전하는조리개는광원로에위치하거나광학장치의영상로에위치하고있다. Lava C.O.S. 는측면조리개가영상로에위치하는 3차원이미지시스템의원리를보여주고있다. 24 광학장치는대상을보는렌즈들을의미하며회전체는측면조리개의요소로회전하는요소이다. 화상처리장치는화상처리과정을수행하는하드웨어를뜻한다. 화상처리장치와회전체는서로조절을위해피드백과정을가진다. Fig. 6는현재 Lava C.O.S. 를구동하는이론인활성파면샘플링을보여주는것으로써, 광파면샘플링을통해초점밖의면을측정하는개념과물체의포인트를조합하는것을보여준다. 이시스템은크게렌즈 (140), 회전하는조리개요소 (160), 상평면 (18 A) 으로이루어져있다. R은측면 pupil을따라원모양으로회전했을때그직경을말하며 d는상평면에생긴초점밖의포인트의이미지를따라그은원의직경이다. 24 하나의조리개는다른물체부분으로부터의이미지들의겹침을피할수있다. 따라서회전하는조리개는여러조리개위치에서얻어진이미지를얻을수있게하며이것은다른시점에있는여러개의카메라에의해해석될수있다. 그러한다양한시점은측정민감도를증가시킬수있다. 활성화로조절된조리개의운동은이미지불균형을조절하여이동탐지의정확성을증가시킬수있게한다. Ray trace at aperture position #1 x y Focal Plane z 8A Out of Focus Point Lens, 140 R Off Axis Exit Pupil, 160A #1 at time t 18A d (X,Y,Z) 0 60 Ray trace at aperture position #2 #2 at time t+δt (X,Y) Image Plane Δz (-) (+) L (-) (+) Z f Z 0 Fig. 6. Lava C.O.S. 의 Active wavefront sampling principle J Dent Rehabil Appl Sci 2015;31(2):112-25

8 Review of recent developments for intra-oral scanners 이미지의가공을위해부분화된상호상관을적용하 여프레임사이에이미지불균형을줄여나갈수있다. 자기상관보다상호상관을사용하는것은장점이많은 데, 자기상관의주요문제점을해결하여소음수준을 낮출수있고, 초점을보다잘잡을수있으며방향적으 로애매한부분을없앨수있다. 25 이러한장점들은노이 즈보다신호의비율을높이고더높은공간적, 심도적 인해상도를가지게하고불확실하게실행할확률을줄 인다. 조리개의움직임은 CCD 요소에다른조리개위 치에있는하나의노출된이미지를저장할수있게해준 다. 부분화된자기상관의작용으로상평면 (18 A) 에최 소두개이상의이미지가기록된다 구강스캐너의적용분야와그분야에서의정확성기술 현재구강스캐너가치과영역에서적용되고있는분 야와그분야에서의적용정도를분석하였다. (1) 수복 Pubmed 와구글스칼라에검색어 : 디지털인상지르 코니아, 디지털인상전부도재관 의검색어로최근 5 년 간논문을검색한후, 관련성이있는데이터만을추출 하여기술하였다. 1 구강스캐너의정확성, 기존의전통방식과의비교 An 등 26 은세라믹크라운제작을위하여준비한하나 의상악중절치를이용하여 30 개의동일한모형을제작 하였다. 10 개씩군을나누었으며각각은 (1) 폴리비닐실 록산인상재로 stone die 를만들어서제작, (2) itero 시 스템으로폴리우레탄 die 를만들어서제작, (3) itero 시 스템으로 die 를만들지않고제작 ( 스테레오리소그라피 파일 ) 한다음변연의차이를관찰하였다. 그결과, (1) µm (2) µm (3) µm 으로측정되었다. 결론적으로, 전통적인방법이 itero 시스템으로인상을 떠서제작한것보다더정확하였으며그결과는유의미 한차이를보였다. 하지만어떤절대적인임상적으로받 아들일수있는변연의차이의양이정해져있지는않으 며세방법모두임상적으로받아들일수있었다. Ng 등 27 에서는 17 번치아의세라믹크라운 30 개를대 상으로 15 개씩나누어서실험하였으며각각은 (1) 전통 적인상방법 ( 폴리비닐실록산인상재 ) 로인상뜬다음제작, (2) 디지털인상방법 : Lava C.O.S. 로인상뜬다음제작하였다. 그다음완성된 30개를근심, 원심, 협측, 구개측, 근심협측, 근심설측, 원심협측, 원심설측, 즉총 8면을보아서 240면의변연의차이를조사하였다. 그결과, (1) 74 µm, (2) 48 µm으로조사되었다. 즉전통적인인상방법이디지털인상방법보다부정확했다. 하지만두방법모두임상적으로양호했다. Lee 등에서는 28 3shape A/S을사용한실험으로각각은 (1) 러버인상재를이용하여만든 working model을스캔하여만든모형, (2) 스캐너를이용하여디지털인상체득후만든디지털모형에서변연의차이를측정하였다. 그결과 (1) 28.9 µm, (2) µm으로조사되었다. 즉 working model로얻은데이터가디지털인상방법의데이터보다정확했고유의미한차이를보였으나두개모두임상적으로받아들일수있었다. Keul 등 29 은 itero 스캐너를이용하여 4-unit 가공의치의변연적합도를분석하였다. 12개는스캔하였고 12개는전통적인인상방법을사용하였다. 그리고이것을가지고각각금속와지르코니아를사용하여다음과같이모형을만들었다 : (1) 전통적인인상방법을사용하여만든금속, (2) 디지털인상방법을사용하여만든금속, (3) 전통적인인상방법을사용하여만든지르코니아, (4) 디지털인상방법을사용하여만든지르코니아였다. 변연의적합도를측정한결과 (1) µm (2) µm (3) µm (4) µm로조사되었으며디지털인상방법이전통적인방식보다더정확하였다. 또한금속으로만든것이지르코니아보다더정확하였으나, 모든인상방법과사용한재료는다임상적으로받아들일수있을정도의정확성을보여주었다. Syrek 등 30 은 (1) Lava C.O.S. 와 (2) 전통적인인상방법 ( 실리콘 ) 으로인상을뜬전부도재관변연적합도를비교하였다. 그결과, (1) 49 µm, (2) 71 µm로측정되었다. 즉, 실리콘을사용한기존의인상방법보다디지털인상방법이유의미하게변연의적합도가더좋았다. 하지만두방법모두임상적으로받아들일수있다. Silva 등 31 은 (1) 전통적인인상방법 12개 ( 폴리에스테르 ) 와 (2) Lava C.O.S. 마스터모델 12개로 4개의지르코니아고정성치아보철물을만들고변연의적합도와치내의적합도를측정하여비교하였다. 그결과변연의적합도는 (1) µm (2) µm, 치내의적합도는 (1) µm (2) µm였다. 따라서유의미하게디 J Dent Rehabil Appl Sci 2015;31(2):

9 Choi JH, Lim YJ, Lee WJ, Han JS, Lee SP 지털인상방법을한것이더정확했지만둘다임상적으로문제가없었다. 이상의연구결과를종합해보면, An 등 26 과 Lee 등 28 은기존의방식이더정확하게나온반면, 나머지연구들에서는구강스캐너를사용한디지털인상방법이더정확한보철물을만들었다. 그러나, 실험에서측정한측정값들은모두달랐음에도불구하고모두임상적으로허용가능한범위내에있었다. 2 재료에따른차이 Keul 등 29 에서는지르코니아와금속간의차이를볼수있었다. 같은재료의인상재를사용하더라도 ( 디지털인상방법, 전통적인인상방법 ) 금속이지르코니아보다정확한변연의적합도를보이는결과가있었다. 그리고 Baig 등 32 은 (1) Cercon Y-TZP (yttria-stabilized tetragonal zirconia polycrystals) 와 (2) IPS Empress II (3) 전부금속관의변연의차이를비교하였다. 그결과 (1) 66.4 µm, (2) 36.6 µm, (3) 37.1 µm의차이를보였다. 지르코니아가조금더부정확하기는했지만임상적으로적용하기에는문제가없다는결론을지을수있었다. 금관과비교하였을때는, Wang 등 33 는 35개구치부 CAD/ CAM 지르코니아를 dental wings s50로제작후, 30개의금관과비교하여 2년간확인하였다. 그결과, 균열여부, 심미성, 접촉마모에관해문제가없었으나대합치만조금마모되어있었다. 즉, 지르코니아로수복해도금관과같이 2년은파절도없고접촉마모의문제도없지만, 대합치마모의문제가있기때문에상악, 하악양쪽을다하는것이낫다는결론을내렸다. Scotti 등 34 은 Lava C.O.S. 로제작한구치부와전치부에있는 37개의단일지르코니아도재관을대상으로, 변연의차이, mid-axial 부위, axio-occlusal 부위, 중심교합부위에서모형과금관내부사이의거리를측정하였다. 그결과각각 µm, 112,25 µm, µm, µm였으며, 구치부와전치부간에유의미한차이를보이지않았다. 즉, Lava C.O.S를사용하면구치부나전치부나상관없이모두다임상적으로받아들일수있는범위내로만들어진다는것을의미한다. 또비슷한연구로 Tamim 등 35 에따르면 50명의디지털인상방법으로제작한금속도재관에서모형과금관내부사이의거리를조사해본결과, 변연의적합도 : 46 µm, mid-axial 부위 : 94 µm, 중심교합부위 : 185 µm을보였다. 이역시모두임상적으로받아들일수있는범위내에있었다. 3스캐너에따른차이 Brawek 등 36 은 (1) Lava C.O.S. 와 (2) CEREC AC와의비교하기위해 14개의지르코니아로변연의차이, mid-axial, axio-occlusal, 중심교합부위에서모형과도재관내부사이의거리를측정하였다. 그결과, 변연의차이는 (1) 51 µm (2) 83 µm였고, mid-axial 부위는 (1) 130 µm, (2) 128 µm로기록되었으며, axio-occlusal 부위는 (1) 178 µm (2) 230 µm을보였고, 중심교합은 (1) 181 µm (2) 297 µm로측정되었다. 즉, Lava C.O.S. 가 CEREC AC보다더적합함을보여주며유의미한차이를보였으나두방법모두임상적으로만족스러운결과를보였다. Seelbach 등 37 은 (1) Lava C.O.S., (2) CEREC, (3) itero, (4) 전통적인인상방법 ( 실리콘 ) 으로전부도재관제작후변연의적합도와치내의적합도를각각측정하였다. 변연의적합도는 (1) 48 µm (2) 30 µm (3) 41 µm (4) 44 µm였고, 치내의적합도는 (1) 29 µm (2) 88 µm (3) 50 µm (4) 49 µm였다. 즉변연의적합도는 CEREC이가장정확하였고, 치내의적합도는 Lava C.O.S. 가제일정확하였다. 3가지구강스캐너와전통적인인상방법모두고정된보철수복에적합하였다. 4 Margin preparation에대한차이 Re 등 38 은인공치아 20개를 chamfer finish line 과 rounded shoulder finish line으로준비한후 Lava C.O.S. 로스캔하였다. 그결과둘은전혀유의미한차이가없었다. Souza 등 39 은 (1) tilted chamfer, (2) large chamfer, (3) rounded shoulder finish line으로준비한후변연의적합도과치내의적합도를측정하였다. 그결과, 변연의적합도는 (1) µm, (2) µm, (3) µm이었고, 치내의적합도는 (1) µm, (2) µm, (3) µm로측정되었다. 즉, 변연의적합도의정확성은 rounded shoulder가높았고, 치내의적합도의정확성은 large chamfer가높았다. Renne 등 40 은 Preparation의정확성에따른변연의차이을조사하였다. 62명의임상가가 1시간씩책보고프랩을하여서 (1) Ideal preparations (25명) (2) Considered fair (34명) (3) Categorized as poor (15명) 로책에설명된올바른 preparation에대한지침을읽고판단을내렸다. 그후변연의차이를조사한결과 (1) 38.5 µm (2) 58.3 µm (3) 90.1 µm으로측정되었다. 즉, preparation이정확할수록디지털인상방법이잘이루어지고최종보철물의변연의차이가더줄어드는것을알수있었다. 120 J Dent Rehabil Appl Sci 2015;31(2):112-25

10 Review of recent developments for intra-oral scanners (2) 임플란트 Pubmed와구글스칼라에검색어 : 디지털인상임플란트 의검색어로최근 5년간논문을검색한후, 관련성이있는데이터만을추출하여기술하였다. 1 임플란트인공치근을심은후인상의정확성 Giménez 등 41 은 itero 스캐너로작동자의경험, 식립각도, 깊이에따른스캐닝의정확도차이를알아보았다. 그결과, 작동자의경험은별로상관이없었고각도가커진다고해서디지털인상방법의정확성이줄어들지않았다. 또한깊이는깊다고해서정확성이줄어들지않았다. 다만스캔하는구역이커지면커질수록정확성은줄어들었다. Abdel-Azim 등 42 은 single unit과전악임플란트모형에대해각각디지털인상 / 기공방법과전통적인인상 / 기공방법을시행하여협 / 설측에서변연적합도를측정, 비교하였다. 단일임플란트시전통적인인상 / 기공방법은 24.1 µm, 디지털인상 / 기공방법은 µm로기존의전통방식이더변연의적합도가잘맞았다. 전악임플란트모형시전통적인인상 / 기공방법은 µm, 디지털인상 / 기공방법은 µm로디지털인상 / 기공방법결과의변연의적합도가더높았다. 하지만 single unit과전악임플란트모형에서전통적방법과디지털방법모두임상적으로문제없을정도의변연의적합도를보였다. Papaspyridakos 등 43 은 2000 개가넘는논문을리뷰해서 76개의연구중 41개는무치악, 35개는부분무치악환자로나누어서연구하였는데, 이중 3개의부분무치악환자에대한논문이디지털인상에대한연구였다고밝혔다. 세연구모두기존의오픈트레이방식보다디지털인상방식이덜정확하다고보고하였다. Stoetzer 등 44 은디지털인상방식을응용하여임플란트식립후인공치근의위치파악을시도하였으며 Cone beam 컴퓨터단층촬영과비교하였을때정확도가떨어지지않는다고하였다. 따라서, x-ray를찍지않고임플란트의위치를알수있는장점이있으므로편리할뿐만아니라, 환자의방사선노출량을줄일수있다고하였다. 2 디지털인상을이용한유도된임플란트수술 Nickenig과 Eitner 45 는 23개의임플란트를수술표본으로사용후식립시계획했던것과실제식립위치를확인하였는데, 4.2도정도차이가나고임플란트 shoulder는 0.9 mm 차이, 임플란트근첨은 0.6 mm 정도차이가났다. 이정도차이는임상적으로허용될만하다고결론지었다. Mandelaris와 Vlk 46 과 Stapleton 47 은디지털인상을이용한임플란트수술에대한케이스를발표하였다. CAD/CAM fabricated, 임플란트유지형고정성치아보철물로서디지털진단인상을제작한후이데이터를 cone beam CT와연계하여정확한진단을내렸다. 또한, 가상계획을통해정확한각도와깊이를계산해내고컴퓨터가유도하여수술견본을만들어심은후, 즉각적인예비보철까지구강스캐너를사용하여진행하였다. 그후과정은전통적인보철물제작과정을사용하였으며보철물에대한이상은없었다. (3) 가철성보철물가철성보철물에대한디지털인상에관련된논문을발견하기가어려웠다. Papaspyridakos 등 43 도 2879개의논문을리뷰하였으나, 무치악환자에대한디지털인상의정확성에관한논문은찾을수없다고밝혔다. 하지만, 최근발표된논문중에서, Ender와 Mehl 48 은전악인상으로, CEREC의디지털인상과전통방식의인상 ( 비닐실록산인상재 ) 을정확성과정밀성의관점에서비교하였다. 그결과, 편위의패턴이전통방식과디지털방식이달랐다. 전통방식으로인상채득한것은전치부와소구치부분은편위가작고원심단부분은 50 µm로가장크고왜곡이있었다. 그에비해디지털방식은제일심한부분은 170 µm 차이가나고, 전체적인모델이시상, 수평축모두다심하게왜곡되어있었다. 하지만 170 µm의차이는임상적으로받아들이지못할오차는아니라고앞선수많은논문에서결론을내렸다. Kattadivil 등 49 은 itero 구강스캐너를통해 Kennedy class II에대한가철성국소의치를완성한첫사례보고를하였다. 이에따르면, 치아와교합을 28번스캔하였고레스트시트, 안내면, 연부조직과상악구개를 25번스캔하였으며대합치또한 28번스캔하여총 81번의스캔을 17분동안진행하였으며, Cadent itero를가지고폴리우레탄주물을만들어가상설계를해당소프트웨어인 SensAble (Sensable Technology Inc. s ) 을통해진행하여레진패턴을 3D 프린팅하였다고보고하였다. 그다음과정은모두전통적인방법으로주조하고제작하였으며최종보철물의적합도는정확하였고환자는불편을느끼지않았다고보고하였다. J Dent Rehabil Appl Sci 2015;31(2):

11 Choi JH, Lim YJ, Lee WJ, Han JS, Lee SP 결론 연구한구강스캐너들은삼각법 (triangulation technique), 공초점레이저현미경 (confocal laser scanner microscopy), 활성파면샘플링 (active wavefront sampling) 등의방식을사용하고있으며, 이방법들을사용하고있 는스캐너들인 CEREC, itero, Lava C.O.S., Trios 의정 확성또한모두실험마다차이는있지만, 모두일정수 준이상의정확성을보이고있으며임상적으로무리가 없을정도의수준에이르렀다. 일부실험에서는디지털 인상이임상적기준을넘었을뿐만아니라, 기존전통방 식으로인상을떠서만든보철물보다더정확하기도하 였다. 그리고그정확성은디지털인상후만들어지는 보철물의재료에따라상이한결과를보였고인상시마 진의모양이나정확성에따라차이를보였다. 이들스캐 너들은치과분야에서간단한단일수복부터교의치, 임 플란트, 고정성치아보철물, 가철성보철물등의제작 과정에도일부사용되고있으며, 최종적으로만들어진 보철물들의적합도, 안정감, 유지가임상적으로문제가 없을정도의수준이라고보고되고있다. 하지만아직디 지털인상방식에대한정확성이나사례보고등이많이 부족하므로, 디지털인상이치과진료에서더큰도움이 되기위해서는수많은추후연구가필요하다. Acknowledgments 본연구는산업통상자원부의재원으로산업핵심기술 개발사업의지원에의하여이루어진것임 ( ). Orcid Young-Jun Lim Jung-Suk Han Seung-Pyo Lee References 1. Ali P, Behrai T. A reliable 3D laser triangulationbased scanner with a new simple but accurate procedure for finding scanner parameters. J Am Sci 2010;6: Kostiukova VV, Riakhovski AN, Ukhanov MM. A Comparative Analysis Of Intra-oral 3d digital scanners for restorative dentistry. Stomatologiia (Mosk) 2014;93: Franca JGDM, Gazziro MA, Saito JH. A 3D scanning system based on laser triangulation and variable field of view. IEEE Intl Conf 2005;1: Wang L, Bo M, Gao J, Ou CS. A Novel double triangulation 3D camera design. IEEE Intl Conf Info 2006: El-Hakim SF, Beraldin JA, Blais F. Comparative evaluation of the performance of passive and active 3D vision systems. Proc. SPIE 2646, Digital Photogrammetry and Remote Sensing 95, 14 (December 1, 1995); doi: / Lombardo V, Marzulli T, Pappalettere C, Sforza P. A time-of-scan laser triangulation technique for distance measurements. Opt Laser Eng 2003;39: Logozzo S, Zanetti EM, Franceschini G, Kilpelä A, Mäkynen A. Recent advances in dental optics - Part I: 3D intra-oral scanners for restorative dentistry. Opt Laser Eng 2014;54: Birnbaum NS, Aaronson HB, Stevens C, Cohen B. 3D digital scanners: a high tech approach to more accurate dental impressions. Inside Dent 2009;5: Thiel F, Pfeiffer J, Fornoff P. Apparatus and method for optical 3D measurement. US Patent 2011/ Thiel E, Pfeiffer J, Fornoff P. Apparatus and method for optical 3D measurement. International Publication WO Patent 2008/ B Schmidt V. 3D dental camera for recording surface structures of a measuring object by means of triangulation. International Publication WO Patent 2010/ A Pawley JB. Handbook of biological confocal microscopy. 3 rd ed. NewYork; Springer; Fellers TJ, Davidson MW. Olympus FluoView Resource Center: introduction to confocal microscopy. Available from: com/theory/confocalintro.html (updated 2014 Oct 15). 14. Babayoff N, Glaser-Inbari I. Imaging a three dimensional structure by confocal focusing an array of light beams. International Publication WO Pat- 122 J Dent Rehabil Appl Sci 2015;31(2):112-25

12 Review of recent developments for intra-oral scanners ent 2000/ Harrison L. Digital impressions competition booming, Available from: index.aspx?sec=sup&sub=rst&pag=dis&itemid= (updated 2014 Oct 15). 16. Babayoff N, Glaser-Inbari I. Method and apparatus for imaging three-dimensional structure. US Patent 2007/ Babayoff N. Method and apparatus for colour imaging a three dimensional structure. US Patent 2010/ A Jacobson B. Taking the headache out of impressions. Dent Today 2007;26:74, Shape. 3Shape TRIOSs digital impression solution. Available from: (updated 2014 Oct 15). 20. Fisker R, Öjelund H, Kjær R, VanDerPoel M, Qazi AA, Hollenbeck KJ. Focus scanning apparatus. International Publication WO Patent 2010/ A Heber S. 3d image reconstruction using active wavefront sampling. Available from: icg.tugraz.at/project_page/project_wavefront/ project_wavefront.htm (updated 2014 Oct 15). 22. Fasbinder DJ. Digital workflow for the Lava C.O.S system. Inside Dent 2009; M ESPE. Lava TM ScanSToptical scanning system. Available from: portal/3m/en_za/3m_espe/dental Manufacturers/Products/DigitalDentistry/Dental Technician/DentalCAD-CAM/DentalScanners (updated 2014 Oct 15). 24. Hart DP, Lammerding J, Rohaly J. 3-D imaging system. US Patent 2004/ A Kondon N. Making the switch to the Lava C.O.S. Available from: News/Making_the_Switch_to_the_Lava_C_O_ S_.aspx (updated 2014 Oct 15). 26. An S, Kim S, Choi H, Lee JH, Moon HS. Evaluating the marginal fit of zirconia copings with digital impressions with an intra-oral digital scanner. J Prosthet Dent 2014;112: Ng J, Ruse D, Wyatt C. A comparison of the marginal fit of crowns fabricated with digital and conventional methods. J Prosthet Dent 2014;112: Lee WS, Kim WC, Kim HY, Kim WT, Kim JH. Evaluation of different approaches for using a laser scanner in digitization of dental impressions. J Adv Prosthodont 2014;6: Keul C, Stawarczyk B, Erdelt KJ, Beuer F, Edelhoff D, Güth JF. Fit of 4-unit FDPs made of zirconia and CoCr-alloy after chairside and labside digitalization-a laboratory study. Dent Mater 2014;30: Syrek A, Reich G, Ranftl D, Klein C, Cerny B, Brodesser J. Clinical evaluation of all-ceramic crowns fabricated from intra-oral digital impressions based on the principle of active wavefront sampling. J Dent 2010;38: Almeida e Silva JS, Erdelt K, Edelhoff D, Araújo É, Stimmelmayr M, Vieira LC, Güth JF. Marginal and internal fit of four-unit zirconia fixed dental prostheses based on digital and conventional impression techniques. Clin Oral Investig 2014;18: Baig MR, Tan KB, Nicholls JI. Evaluation of the marginal fit of a zirconia ceramic computer-aided machined (CAM) crown system. J Prosthet Dent 2010;104: Wang YG, Xing YX, Sun YC, Zhao YJ, Lü PJ, Wang Y. Preliminary evaluation of clinical effect of computer aided design and computer aided manufacture zirconia crown. Zhonghua Kou Qiang Yi Xue Za Zhi 2013;48: Scotti R, Cardelli P, Baldissara P, Monaco C. WITHDRAWN: Clinical fitting of CAD/CAM zirconia single crowns generated from digital intraoral impressions based on active wavefront sampling. J Dent 2011 Oct 17.[Epub ahead of print] 35. Tamim H, Skjerven H, Ekfeldt A, Rønold HJ. Clinical evaluation of CAD/CAM metal-ceramic posterior crowns fabricated from intra-oral digital impressions. Int J Prosthodont 2014;27: Brawek PK, Wolfart S, Endres L, Kirsten A, Reich S. The clinical accuracy of single crowns exclusively fabricated by digital workflow-the comparison of two systems. Clin Oral Investig 2013;17: Seelbach P, Brueckel C, Wöstmann B. Accuracy of digital and conventional impression techniques and workflow. Clin Oral Investig 2013;17: J Dent Rehabil Appl Sci 2015;31(2):

13 Choi JH, Lim YJ, Lee WJ, Han JS, Lee SP 38. Re D, Cerutti F, Augusti G, Cerutti A, Augusti D. Comparison of marginal fit of Lava CAD/CAM crown-copings with two finish lines. Int J Esthet Dent 2014;9: Souza RO, Özcan M, Pavanelli CA, Buso L, Lombardo GH, Michida SM, Mesquita AM, Bottino MA. Marginal and internal discrepancies related to margin design of ceramic crowns fabricated by a CAD/CAM system. J Prosthodont 2012;21: Renne W, McGill ST, Forshee KV, DeFee MR, Mennito AS. Predicting marginal fit of CAD/ CAM crowns based on the presence or absence of common preparation errors. J Prosthet Dent 2012; 108: Giménez, B, Özcan, M, Martínez-Rus, F, Pradíes G. Accuracy of a digital impression system based on parallel confocal laser technology for implants with consideration of operator experience and implant angulation and depth. Int J Oral Maxillofac Implants 2014;29: Abdel-Azim T, Zandinejad A, Elathamna E, Lin W, Morton D. The influence of digital fabrication options on the accuracy of dental implant-based single units and complete-arch frameworks. Int J Oral Maxillofac Implants 2014;29: Papaspyridakos P, Chen CJ, Gallucci GO, Doukoudakis A, Weber HP, Chronopoulos V. Accuracy of implant impressions for partially and completely edentulous patients: a systematic review. Int J Oral Maxillofac Implants 2014;29: Stoetzer M, Wagner ME, Wenzel D, Lindhorst D, Gellrich NC, von See C. Nonradiological method for 3-dimensional implant position assessment using an intra-oral scan: new method for postoperative implant control. Implant Dent 2014; 23: Nickenig HJ, Eitner S. An alternative method to match planned and achieved positions of implants, after virtual planning using cone-beam CT data and surgical guide templates-a method reducing patient radiation exposure (part I). J Craniomaxillofac Surg 2010;38: Mandelaris GA, Vlk SD. Guided implant surgery with placement of a presurgical CAD/CAM patient-specific abutment and provisional in the esthetic zone. Compend Contin Educ Dent 2014; 35: Stapleton BM, Lin WS, Ntounis A, Harris BT, Morton D. Application of digital diagnostic impression, virtual planning, and computer-guided implant surgery for a CAD/CAM-fabricated, implantsupported fixed dental prosthesis: a clinical report. J Prosthet Dent 2014;112: Ender A, Mehl A. Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. J Prosthet Dent 2013;109: Kattadiyil MT, Mursic Z, AlRumaih H, Goodacre CJ. Intra-oral scanning of hard and soft tissues for partial removable dental prosthesis fabrication. J Prosthet Dent 2014;112: J Dent Rehabil Appl Sci 2015;31(2):112-25

14 Review Article 현재존재하는구강스캐너에대한고찰 최종훈 1, 임영준 2, 이원진 3, 한중석 2, 이승표 4 * 1 서울대학교치의학대학원 2 서울대학교치의학대학원치과보철학교실 3 서울대학교치의학대학원구강악안면방사선학교실 4 서울대학교치의학대학원구강해부학교실 구강내의모습을재현하는복제모델을만드는것은치과진료에서가장중요한과정이며정확성과효율성이동시에만족되어야하는과정이다. 현재기술이발전함에따라치과진료에서도디지털화가이루어지고있다. 이러한것을가능하게하는가장중요한작업중하나가바로구강내의모습을 3차원적으로재구성하는디지털화이다. CAD/CAM 시스템의 3가지성분 1 data capture component (digitizers), 2 design component (CAD software), 3 manufacturing component (CAM) 중가장기본이되며뒤의과정에막대한영향을끼치는것이 data capture component 즉구강스캐너이다. 이논문은 Pubmed와 Google Scholar에서최근 5년전연구논문들을기초로하여, 각각의스캐너의구동원리와스캐너들간의정확성, 현재구강스캐너가치과영역에서적용되고있는분야와그정도를분석하였다. ( 구강회복응용과학지 2015;31(2):112-25) 주요어 : 구강스캐너 ; 디지털인상 ; 3-d 이미지화 * 교신저자 : 이승표 ( ) 서울특별시종로구연건동 28 번지서울대학교치의학대학원구강해부학교실 Tel: Fax: orana9@snu.ac.kr 접수일 : 2015 년 2 월 12 일 수정일 : 2015 년 5 월 15 일 채택일 : 2015 년 5 월 31 일 125

ePapyrus PDF Document

ePapyrus PDF Document .130-812 103-18 02-2253-2800 02-2253-2809 kdtech@chol.com 370 2013 10 1 1977 11 20 Korean Dental Technologist Association News, 45 8 30. 6 27 2 2. ( ),,,,,,,,,, 13.,,., ( ) 40. 9 27 45 9 27 45 6. (1,272

More information

Digital Abutment

Digital Abutment Digital Abutment P25 Digital Abutment Product Catalog 기성 Abutment 보다진화된, Customized Abutment 보다편리한, Digital Abutment! 절치 소구치 대구치 CEJ Cemento Enamel junction Digital Abutment 는 CEJ 를기준으로생성되어보철물의심미성이우수하며,

More information

CASE REPORT 구강스캐너를이용한임플란트보철물제작증례 강병길 김희중 정재헌 * 조선대학교치과대학치과보철학교실 Implant prosthesis using intraoral scanne

CASE REPORT   구강스캐너를이용한임플란트보철물제작증례 강병길 김희중 정재헌 * 조선대학교치과대학치과보철학교실 Implant prosthesis using intraoral scanne CASE REPORT http://dx.doi.org/10.4047/jkap.2015.53.3.256 강병길 김희중 정재헌 * 조선대학교치과대학치과보철학교실 Implant prosthesis using intraoral scanner: Case Report Byeong-Gil Kang, Hee-Jung Kim, Chae-Heon Chung* Department

More information

ORIGINAL ARTICLE CAD-CAM 을이용하여디자인한금관의치과직업군에따른형태비교 김태현 1 김종은 1 이아름 1,2 박영범 1,2 * 1 연세대학교치과대학치과보철학교실 2 연세대학

ORIGINAL ARTICLE   CAD-CAM 을이용하여디자인한금관의치과직업군에따른형태비교 김태현 1 김종은 1 이아름 1,2 박영범 1,2 * 1 연세대학교치과대학치과보철학교실 2 연세대학 ORIGINAL ARTICLE http://dx.doi.org/0.4047/jkap.06.54.3.34 김태현 김종은 이아름, 박영범, * 연세대학교치과대학치과보철학교실 연세대학교치과대학구강과학연구소, BK 플러스통합구강생명과학사업 Comparison of crown designs of different dental occupational groups, using

More information

ISSN Implantology 2014; 18(2): 140~146 디지털 3 차원영상을이용한무피판절개임플란트식립 : 증례보고 백장현 경희대학교치의학전문대학원치과보철학교실 Flapless Implant Placement with Digital Thr

ISSN Implantology 2014; 18(2): 140~146 디지털 3 차원영상을이용한무피판절개임플란트식립 : 증례보고 백장현 경희대학교치의학전문대학원치과보철학교실 Flapless Implant Placement with Digital Thr ISSN 1229-5418 Implantology 2014; 18(2): 140~146 디지털 3 차원영상을이용한무피판절개임플란트식립 : 증례보고 백장현 경희대학교치의학전문대학원치과보철학교실 Flapless Implant Placement with Digital Three-dimensional Imaging and Planning System: A Case

More information

untitled

untitled 전방향카메라와자율이동로봇 2006. 12. 7. 특허청전기전자심사본부유비쿼터스심사팀 장기정 전방향카메라와자율이동로봇 1 Omnidirectional Cameras 전방향카메라와자율이동로봇 2 With Fisheye Lens 전방향카메라와자율이동로봇 3 With Multiple Cameras 전방향카메라와자율이동로봇 4 With Mirrors 전방향카메라와자율이동로봇

More information

09권오설_ok.hwp

09권오설_ok.hwp (JBE Vol. 19, No. 5, September 2014) (Regular Paper) 19 5, 2014 9 (JBE Vol. 19, No. 5, September 2014) http://dx.doi.org/10.5909/jbe.2014.19.5.656 ISSN 2287-9137 (Online) ISSN 1226-7953 (Print) a) Reduction

More information

ORIGINAL ARTICLE 디지털구강스캐너로모형없이제작한전부지르코니아수복물의변연및내면적합도평가 이종원 1 박지만 2 * 1 이대목동병원치과보철과, 2 관악서울대학교치과병원치과보철과, 치

ORIGINAL ARTICLE   디지털구강스캐너로모형없이제작한전부지르코니아수복물의변연및내면적합도평가 이종원 1 박지만 2 * 1 이대목동병원치과보철과, 2 관악서울대학교치과병원치과보철과, 치 ORIGINAL ARTICLE http://dx.doi.org/10.4047/jkap.2016.54.3.210 이종원 1 박지만 2 * 1 이대목동병원치과보철과, 2 관악서울대학교치과병원치과보철과, 치학연구소 Evaluation of marginal and internal gap under model-free monolithic zirconia restoration

More information

(

( 317 318 319 320 1 3 5 5 5 5 2 321 : 1.,,,,, 06 2. X-ray beam penetration (density) (contrast) 03 3. patch coating, precipitation, flaking 03 4. centering 03 5. Esophagus, cardia, fundus, body, angle, antrum,

More information

저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할

저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할 저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할수없습니다. 변경금지. 귀하는이저작물을개작, 변형또는가공할수없습니다. 귀하는, 이저작물의재이용이나배포의경우,

More information

<283732372D3733312920B4D9C3CAC1A120BCD2C7C1C6AEC4DCC5C3C6AEB7BBC1EEC0C720B3EBBEC8C0C720BDC3B7C2BAB8C1A4BFA120B4EBC7D120C0AFBFEBBCBA20C6F2B0A1283035292E687770>

<283732372D3733312920B4D9C3CAC1A120BCD2C7C1C6AEC4DCC5C3C6AEB7BBC1EEC0C720B3EBBEC8C0C720BDC3B7C2BAB8C1A4BFA120B4EBC7D120C0AFBFEBBCBA20C6F2B0A1283035292E687770> 대한안과학회지 제 49 권 제 5 호 2008 J Korean Ophthalmol Soc 49(5):727-731, 2008 DOI : 10.3341/jkos.2008.49.5.727 다초점 소프트콘택트렌즈의 노안의 시력보정에 대한 유용성 평가 김현경 1 김효명 2 정성근 1 가톨릭대학교 의과대학 성모병원 안과학교실 1, 고려대학교 의과대학 안암병원 안과학교실

More information

ISSN Implantology 2016; 20(2): 66~71 시멘트를사용하지않는새로운유지형태를가진임플란트고정성보철물 : 증례보고 이현종, 백장현 경희대학교치의학전문대학원보철학교실 A New Retentive Type of Fixed Implant

ISSN Implantology 2016; 20(2): 66~71 시멘트를사용하지않는새로운유지형태를가진임플란트고정성보철물 : 증례보고 이현종, 백장현 경희대학교치의학전문대학원보철학교실 A New Retentive Type of Fixed Implant ISSN 1229-5418 Implantology 2016; 20(2): 66~71 시멘트를사용하지않는새로운유지형태를가진임플란트고정성보철물 : 증례보고 이현종, 백장현 경희대학교치의학전문대학원보철학교실 A New Retentive Type of Fixed Implant Prosthesis without Cement: A Case Report Hyeonjong

More information

example code are examined in this stage The low pressure pressurizer reactor trip module of the Plant Protection System was programmed as subject for

example code are examined in this stage The low pressure pressurizer reactor trip module of the Plant Protection System was programmed as subject for 2003 Development of the Software Generation Method using Model Driven Software Engineering Tool,,,,, Hoon-Seon Chang, Jae-Cheon Jung, Jae-Hack Kim Hee-Hwan Han, Do-Yeon Kim, Young-Woo Chang Wang Sik, Moon

More information

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. vol. 29, no. 10, Oct ,,. 0.5 %.., cm mm FR4 (ε r =4.4)

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. vol. 29, no. 10, Oct ,,. 0.5 %.., cm mm FR4 (ε r =4.4) THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 2018 Oct.; 29(10), 799 804. http://dx.doi.org/10.5515/kjkiees.2018.29.10.799 ISSN 1226-3133 (Print) ISSN 2288-226X (Online) Method

More information

untitled

untitled .130-812 103-18 02-2253-2800 02-2253-2809 kdtech@chol.com www.kdtech.or.kr 373 2014 1 1 1977 11 20 Korean Dental Technologist Association News 2013 12 31 2,. CAD CAM. ( ) 2011 4,,,,,,,, 6 5 ( ) 2011 7

More information

하단에서평균 1.51 mm 위치변위가생겼다. 발생하는각도오차는평균 4.9 이었으며, 최대오차값은 15.3 이었다. 4-9 각도오차 15 는인접한해부학적구조물에치명적인손상을줄수있는오차값이다. 이들문헌에보고된수술가이드는대부분수작업으로방사선가이드를제작하고이를수술가이드로전환하

하단에서평균 1.51 mm 위치변위가생겼다. 발생하는각도오차는평균 4.9 이었으며, 최대오차값은 15.3 이었다. 4-9 각도오차 15 는인접한해부학적구조물에치명적인손상을줄수있는오차값이다. 이들문헌에보고된수술가이드는대부분수작업으로방사선가이드를제작하고이를수술가이드로전환하 ORIGINAL ARTICLE http://dx.doi.org/10.4047/jkap.2015.53.2.111 정승미 1 * 방정환 1 황찬현 1 강세하 1 최병호 2 방일흠 2 전형태 3 안성훈 3 연세대학교원주의대 1 치과보철과, 2 구강악안면외과, 3 부산디오임플란트연구소 Accuracy assessment of implant placement using

More information

I 서론 two-piece 지르코니아어버트먼트를사용하여상악단일임플란트보철을제작하였으며, 심미적, 기능적으로만족할만한결과를얻었기에이를보고한다. 임플란트고정성보철물은임플란트주위연조직이인접연조직과자연스럽게조화를이루고보철물은자연치의적절한외형을재현해야한다 1. 타이타늄임플란트어

I 서론 two-piece 지르코니아어버트먼트를사용하여상악단일임플란트보철을제작하였으며, 심미적, 기능적으로만족할만한결과를얻었기에이를보고한다. 임플란트고정성보철물은임플란트주위연조직이인접연조직과자연스럽게조화를이루고보철물은자연치의적절한외형을재현해야한다 1. 타이타늄임플란트어 ISSN 1229-5418 Implantology 2013; 17(2): 98~107 CAD/CAM Two-Piece Zirconia Abutment 를이용한상악전치부의단일임플란트식립과수복증례 이도연, 박수정, 이성복, 이석원, 안수진강동경희대학교병원치과병원생체재료보철학교실 CAD/CAM Two-Piece Zirconia Abutments Used for Anterior

More information

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Dec.; 27(12),

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Dec.; 27(12), THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 2016 Dec.; 27(12), 1036 1043. http://dx.doi.org/10.5515/kjkiees.2016.27.12.1036 ISSN 1226-3133 (Print) ISSN 2288-226X (Online)

More information

ISSN eissn Case Report Full mouth rehabilitation utilizing implant-assisted removab

ISSN eissn Case Report Full mouth rehabilitation utilizing implant-assisted removab http://dx.doi.org/10.14368/jdras.2016.32.3.214 ISSN 2384-4353 eissn 2384-4272 Case Report Full mouth rehabilitation utilizing implant-assisted removable partial denture with a canted occlusal plane: a

More information

Kbcs002.hwp

Kbcs002.hwp Does Real-time Compound Imaging Improve Evaluation of reast Cancer Compared to Conventional Sonography? o Kyoung Seo, M.D., Yu Whan Oh, M.D., Kyu Ran Cho, M.D., Young Hen Lee, M.D., Hyung Joon Noh, M.D.,

More information

139~144 ¿À°ø¾àħ

139~144 ¿À°ø¾àħ 2 139 DOI : 10.3831/KPI.2010.13.2.139 2 Received : 10. 04. 08 Revised : 10. 04. 26 Two Case Report on Wrist Ganglion Treated with Scolopendrid Pharmacopuncture Accepted : 10. 05. 04 Key Words: Wrist Ganglion,

More information

CASE REPORT 디지털장비의중첩기능을이용하여단일체수복물의교합조정을최소화한증례 최창훈 김선재 * 연세대학교강남세브란스병원치과보철과 Superimposition: a simple meth

CASE REPORT   디지털장비의중첩기능을이용하여단일체수복물의교합조정을최소화한증례 최창훈 김선재 * 연세대학교강남세브란스병원치과보철과 Superimposition: a simple meth CSE REPORT http://dx.doi.org/10.4047/jkap.2016.54.3.253 최창훈 김선재 * 연세대학교강남세브란스병원치과보철과 Superimposition: a simple method to minimize occlusal adjustment of monolithic restoration Changhun Choi, Sunjai Kim*

More information

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Jul.; 27(7),

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Jul.; 27(7), THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 16 Jul.; 27(7), 64662. http://dx.doi.org/./kjkiees.16.27.7.646 ISSN 1226-3133 (Print)ISSN 2288-226 (Online) 2D Microwave Image

More information

ABSTRACT QLF Concept and Clinical Implementation Department of Preventive Dentistry and Public Oral Health, College of Dentistry, Yonsei University Baek Il Kim, D.D.S., Ph.D. The leading paradigm of dentistry

More information

09È«¼®¿µ 5~152s

09È«¼®¿µ5~152s Korean Journal of Remote Sensing, Vol.23, No.2, 2007, pp.45~52 Measurement of Backscattering Coefficients of Rice Canopy Using a Ground Polarimetric Scatterometer System Suk-Young Hong*, Jin-Young Hong**,

More information

ISSN eissn Review Article The user-friendly, ergonomic design aspect of recent intr

ISSN eissn Review Article The user-friendly, ergonomic design aspect of recent intr http://dx.doi.org/.368/jdras.2015.31.3.221 ISSN 2384-4353 eissn 2384-4272 Review Article The user-friendly, ergonomic design aspect of recent intraoral scanners: a literature review Hye-Nan Park 1, Won-Hee

More information

저작자표시 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 이저작물을영리목적으로이용할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 변경금지. 귀

저작자표시 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 이저작물을영리목적으로이용할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 변경금지. 귀 저작자표시 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 이저작물을영리목적으로이용할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 변경금지. 귀하는이저작물을개작, 변형또는가공할수없습니다. 귀하는, 이저작물의재이용이나배포의경우, 이저작물에적용된이용허락조건을명확하게나타내어야합니다.

More information

Park JW, Kim KR, Kang HW, Lee KB, Lee DH 다 드 제작 프로그램(DDS-Pro)으로 전송하였다. 프로그 본 증례에서는 최근에 개발된 컴퓨터 단층촬영에 램을 이용하여 악골과 공간 분석 후 임플란트 형태를 입 기반한 임플란트 가이드

Park JW, Kim KR, Kang HW, Lee KB, Lee DH 다 드 제작 프로그램(DDS-Pro)으로 전송하였다. 프로그 본 증례에서는 최근에 개발된 컴퓨터 단층촬영에 램을 이용하여 악골과 공간 분석 후 임플란트 형태를 입 기반한 임플란트 가이드 http://dx.doi.org/10.14368/jdras.2015.31.1.67 ISSN 2233-4084 Case Report Implant placement using a newly developed CT-based guide program and subtractive manufacturing: case reports Jung-Wan Park, Kyung-Rok

More information

08원재호( )

08원재호( ) 30 2 20124 pp. 173~180 Non-Metric Digital Camera Lens Calibration Using Ground Control Points 1) 2) 3) Abstract The most recent, 80 mega pixels digital camera appeared through the development of digital

More information

<313920C0CCB1E2BFF82E687770>

<313920C0CCB1E2BFF82E687770> 韓 國 電 磁 波 學 會 論 文 誌 第 19 卷 第 8 號 2008 年 8 月 論 文 2008-19-8-19 K 대역 브릭형 능동 송수신 모듈의 설계 및 제작 A Design and Fabrication of the Brick Transmit/Receive Module for K Band 이 기 원 문 주 영 윤 상 원 Ki-Won Lee Ju-Young Moon

More information

High Resolution Disparity Map Generation Using TOF Depth Camera In this paper, we propose a high-resolution disparity map generation method using a lo

High Resolution Disparity Map Generation Using TOF Depth Camera In this paper, we propose a high-resolution disparity map generation method using a lo High Resolution Disparity Map Generation Using TOF Depth Camera In this paper, we propose a high-resolution disparity map generation method using a low-resolution Time-Of- Flight (TOF) depth camera and

More information

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Mar.; 28(3),

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Mar.; 28(3), THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 2017 Mar.; 28(3), 163 169. http://dx.doi.org/10.5515/kjkiees.2017.28.3.163 ISSN 1226-3133 (Print) ISSN 2288-226X (Online) PCB

More information

<372E20B9DABFB5BBF32E687770>

<372E20B9DABFB5BBF32E687770> 대한치주과학회지 : Vol. 37, No. 2(Suppl.), 2007 New maxillary anterior ridge classification according to ideal implant restorative position determined by CAT Young-Sang Park 1, Sang-Choon Cho 2, Kyoung-Nam Kim

More information

untitled

untitled J Dent Hyg Sci Vol. 15, No. 2, 2015, pp.226-231 http://dx.doi.org/10.17135/jdhs.2015.15.2.226 RESEARCH ARTICLE 세가지방식의스캐너종류에따른모형정확도평가 이재준ㆍ박진영ㆍ배소연ㆍ전진훈ㆍ김지환ㆍ김웅철 고려대학교일반대학원보건과학과치의기공전공 Evaluation of the Model

More information

42(2)-06(이효경).fm

42(2)-06(이효경).fm Oral Biology Research, 2018; June 30, 42(2):97-103 Copyright c 2018, Oral Biology Research Institute DOI: 10.21851/obr.42.02.201806.97 Case Report ORAL BIOLOGY RESEARCH Rehabilitation of mandibular partial

More information

µµÅ¥¸àÆ®1

µµÅ¥¸àÆ®1 Quintessence of Dental Technology Vol. 24, No. 6, JUNE 1999 6/99 Journal for Dental Technicians & Prosthodontists Quintessence of Dental Technology QDT 3 9 16 26 36 52 CONTENTS June volume24 No. 6 1999

More information

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Jun.; 27(6),

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Jun.; 27(6), THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 2016 Jun.; 27(6), 495 503. http://dx.doi.org/10.5515/kjkiees.2016.27.6.495 ISSN 1226-3133 (Print) ISSN 2288-226X (Online) Design

More information

의료기기품목별시장환경분석에 관한연구 연세대학교보건환경대학원 의공학전공 김수연

의료기기품목별시장환경분석에 관한연구 연세대학교보건환경대학원 의공학전공 김수연 의료기기품목별시장환경분석에 관한연구 연세대학교보건환경대학원 의공학전공 김수연 의료기기품목별시장환경분석에 관한연구 - i - - ii - - iii - - iv - - v - - vi - 국문요약 의료기기품목별시장환경분석에관한연구 - vii - 제 1 장서론 1.1. 연구의배경및필요성 1.2. 연구내용및방법 표 1. 국내의료기기생산및수출입실적분석방법 - 각연도에생산,

More information

03-16-김용일.indd

03-16-김용일.indd Journal of the Korean Society of Surveying, Geodesy, Photogrammetry and Cartography Vol. 33, No. 2, 95-101, 2015 http://dx.doi.org/10.7848/ksgpc.2015.33.2.95 ISSN 1598-4850(Print) ISSN 2288-260X(Online)

More information

CASE REPORT 임플란트지지서베이드금관을이용한하악의가철성국소의치수복증례 박재호 민병귀 양홍서 박찬 박상원 * 전남대학교치의학전문대학원보철학교실 Clinical application of

CASE REPORT   임플란트지지서베이드금관을이용한하악의가철성국소의치수복증례 박재호 민병귀 양홍서 박찬 박상원 * 전남대학교치의학전문대학원보철학교실 Clinical application of SE REPORT https://doi.org/10.4047/jkap.2018.56.2.154 박재호 민병귀 양홍서 박찬 박상원 * 전남대학교치의학전문대학원보철학교실 linical application of mandibular removable partial denture using implant-supported surveyed crown: case report

More information

πŸ¿Ãƒ‹3ø˘

πŸ¿Ãƒ‹3ø˘ 2년 전, 독일 쾰른에서 개최된 IDS에 집채만 한 밀링기계를 가지고 손바닥만 한 금속재료를 깎아 일반 보철물 및 임플란트 보철물을 제작 하는 시스템이 세계 치과계에 첫선을 보였다. 독일 뮌헨 근교에 위치한 Zfx-GmbH는 15년의 CAD/CAM 기술을 바탕으로 4년 전인 2007년 모든 형태의 보철물을 어떠한 재료로든 밀링해 만들 수 있는 기능을 보유한

More information

ISSN Implantology 2016; 20(3): 132~141 상악무치악에서단일상지르코니아를이용한임플란트 - 지지고정성보철물수복증례 박성우, 권주현, 장재승, 김선재 강남세브란스치과병원치과보철과 Implant-Supported Fixed Pro

ISSN Implantology 2016; 20(3): 132~141 상악무치악에서단일상지르코니아를이용한임플란트 - 지지고정성보철물수복증례 박성우, 권주현, 장재승, 김선재 강남세브란스치과병원치과보철과 Implant-Supported Fixed Pro ISSN 1229-5418 Implantology 2016; 20(3): 132~141 상악무치악에서단일상지르코니아를이용한임플란트 - 지지고정성보철물수복증례 박성우, 권주현, 장재승, 김선재 강남세브란스치과병원치과보철과 Implant-Supported Fixed Prostheses in Maxillary Edentulous Patient with Monolithic

More information

±èÇ¥³â

±èÇ¥³â 533 Fig. 1. Dead zone. Nine echoes are positioned 2-10 mm below the scan surface with 1 mm distance. All nine echoes are clearly visualized. 534 Fig. 2. Vertical and horizontal measurement. 10 cm distance

More information

???? 1

???? 1 The Korean Journal of Applied Statistics (2013) 26(1), 201 208 DOI: http://dx.doi.org/10.5351/kjas.2013.26.1.201 A Note on Model Selection in Mixture Experiments with Process Variables Jung Il Kim a,1

More information

09구자용(489~500)

09구자용(489~500) The Study on the Grid Size Regarding Spatial Interpolation for Local Climate Maps* Cha Yong Ku** Young Ho Shin*** Jae-Won Lee**** Hee-Soo Kim*****.,...,,,, Abstract : Recent global warming and abnormal

More information

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Sep.; 30(9),

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Sep.; 30(9), THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 2019 Sep.; 30(9), 712 717. http://dx.doi.org/10.5515/kjkiees.2019.30.9.712 ISSN 1226-3133 (Print) ISSN 2288-226X (Online) MOS

More information

CASE REPORT 임시치아 double scanning 을이용한전악임플란트수복증례 양동헌 양홍서 박상원 임현필 윤귀덕 방몽숙 * 전남대학교치의학전문대학원치과보철학교실 Full mouth

CASE REPORT   임시치아 double scanning 을이용한전악임플란트수복증례 양동헌 양홍서 박상원 임현필 윤귀덕 방몽숙 * 전남대학교치의학전문대학원치과보철학교실 Full mouth CSE REPORT http://dx.doi.org/10.4047/jkap.2014.52.3.252 양동헌 양홍서 박상원 임현필 윤귀덕 방몽숙 * 전남대학교치의학전문대학원치과보철학교실 Full mouth implant rehabilitation with double scanning of provisional restoration Dong-Hun Yang, Hong-So

More information

07.045~051(D04_신상욱).fm

07.045~051(D04_신상욱).fm J. of Advanced Engineering and Technology Vol. 1, No. 1 (2008) pp. 45-51 f m s p» w Á xá zá Ÿ Á w m œw Image Retrieval Based on Gray Scale Histogram Refinement and Horizontal Edge Features Sang-Uk Shin,

More information

Lumbar spine

Lumbar spine Lumbar spine CT 32 111 DOI : 10.3831/KPI.2010.13.2.111 Lumbar Spine CT 32 Received : 10. 05. 23 Revised : 10. 06. 04 Accepted : 10. 06. 11 Key Words: Disc herniation, CT scan, Clinical analysis The Clinical

More information

45-51 ¹Ú¼ø¸¸

45-51 ¹Ú¼ø¸¸ A Study on the Automation of Classification of Volume Reconstruction for CT Images S.M. Park 1, I.S. Hong 2, D.S. Kim 1, D.Y. Kim 1 1 Dept. of Biomedical Engineering, Yonsei University, 2 Dept. of Radiology,

More information

CASE REPORT Double Scanning 에의한고성능중합체를이용한임플란트지지고정성보철물수복증례 강경환 박진홍 이정열 * 신상완고려대학교구로병원임상치의학연구소 Implant suppor

CASE REPORT   Double Scanning 에의한고성능중합체를이용한임플란트지지고정성보철물수복증례 강경환 박진홍 이정열 * 신상완고려대학교구로병원임상치의학연구소 Implant suppor SE REPORT https://doi.org/10.4047/jkap.2017.55.3.305 강경환 박진홍 이정열 * 신상완고려대학교구로병원임상치의학연구소 Implant supported prosthesis with high performance polymers using a double scanning method Kyeong-Hwan Kang, Jin-Hong

More information

< 목차 > Ⅰ. 연구동기 1 Ⅱ. 연구목적 1 Ⅲ. 연구내용 2 1. 이론적배경 2 (1) 직접제작한물질의기본구조 2 (2) 회절격자의이론적배경 3 (3) X-선회절법-XRD(X-Ray Diffraction) 3 (4) 브래그의법칙 (Bragg`s law) 4 (5)

< 목차 > Ⅰ. 연구동기 1 Ⅱ. 연구목적 1 Ⅲ. 연구내용 2 1. 이론적배경 2 (1) 직접제작한물질의기본구조 2 (2) 회절격자의이론적배경 3 (3) X-선회절법-XRD(X-Ray Diffraction) 3 (4) 브래그의법칙 (Bragg`s law) 4 (5) [ 첨부 4] 작품설명서표지서식 작품번호 1143 LASER 의라우에패턴을통한입체모형의구조분석 출품분야물리출품부문학생 2011. 7. 7 구분성명 출품학생 지도교사 김성현 권채련 김서연 전종술 - 1 - < 목차 > Ⅰ. 연구동기 1 Ⅱ. 연구목적 1 Ⅲ. 연구내용 2 1. 이론적배경 2 (1) 직접제작한물질의기본구조 2 (2) 회절격자의이론적배경 3 (3)

More information

19_9_767.hwp

19_9_767.hwp (Regular Paper) 19 6, 2014 11 (JBE Vol. 19, No. 6, November 2014) http://dx.doi.org/10.5909/jbe.2014.19.6.866 ISSN 2287-9137 (Online) ISSN 1226-7953 (Print) RGB-Depth - a), a), b), a) Real-Virtual Fusion

More information

DBPIA-NURIMEDIA

DBPIA-NURIMEDIA The e-business Studies Volume 17, Number 4, August, 30, 2016:319~332 Received: 2016/07/28, Accepted: 2016/08/28 Revised: 2016/08/27, Published: 2016/08/30 [ABSTRACT] This paper examined what determina

More information

Microsoft Word - Introduction%20to%20guided%20implant%20surgery[1].docx

Microsoft Word - Introduction%20to%20guided%20implant%20surgery[1].docx CYBERMED INC., ONDEMAND3D TECHNOLOGY INC. Introduction to Guided Implant Surgery White Paper Febrary 2010 USA Republic of Korea www.ondemand3d.com What is the guided implant surgery Guided implant surgery란환자의

More information

(Microsoft PowerPoint - \301\24608\260\255 - \261\244\277\370\260\372 \300\347\301\372)

(Microsoft PowerPoint - \301\24608\260\255 - \261\244\277\370\260\372 \300\347\301\372) 게임엔진 제 8 강광원과재질 이대현교수 한국산업기술대학교게임공학과 학습목차 조명모델 광원의색상설정 재질 분산성분의이해 분산재질의구현 경반사성분의이해 경반사재질의구현 조명 (Illumination) 모델 조명모델 광원으로부터공간상의점들까지의조도를계산하는방법. 직접조명과전역조명 직접조명 (direct illumination) 모델 물체표면의점들이장면내의모든광원들로부터직접적으로받는빛만을고려.

More information

ISSN eissn Case Report Full mouth rehabilitation of a patient using monolithic zircon

ISSN eissn Case Report Full mouth rehabilitation of a patient using monolithic zircon https://doi.org/10.14368/jdras.2018.34.3.196 ISSN 2384-4353 eissn 2384-4272 ase Report Full mouth rehabilitation of a patient using monolithic zirconia and dental /M system: a case report Sang-Hoon Lee,

More information

05 목차(페이지 1,2).hwp

05 목차(페이지 1,2).hwp THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 2014 Oct.; 25(10), 10771086. http://dx.doi.org/10.5515/kjkiees.2014.25.10.1077 ISSN 1226-3133 (Print)ISSN 2288-226X (Online)

More information

ISSN Implantology 2016; 20(2): 72~79 양측상악중절치임플란트수복시 Computer Guided Surgical Stent 의이용 이강희 1, 김기덕 2, 김근희 2, 김민유 1, 박원서 1 1 연세해담치과, 2 연세대학교치과

ISSN Implantology 2016; 20(2): 72~79 양측상악중절치임플란트수복시 Computer Guided Surgical Stent 의이용 이강희 1, 김기덕 2, 김근희 2, 김민유 1, 박원서 1 1 연세해담치과, 2 연세대학교치과 ISSN 1229-5418 Implantology 2016; 20(2): 72~79 양측상악중절치임플란트수복시 Computer Guided Surgical Stent 의이용 이강희 1, 김기덕 2, 김근희 2, 김민유 1, 박원서 1 1 연세해담치과, 2 연세대학교치과대학통합진료학과 The Use of Computer Guided Surgical Stent

More information

<32362DC0CCC8F1BCBA2DC7DAB5E5BDBAC4B3B4D720C0DBBEF720B9E6B9FDBFA120B5FBB6F320C7A5C1D820B8F0B5A82E687770>

<32362DC0CCC8F1BCBA2DC7DAB5E5BDBAC4B3B4D720C0DBBEF720B9E6B9FDBFA120B5FBB6F320C7A5C1D820B8F0B5A82E687770> Journal of the Korea Academia-Industrial cooperation Society Vol. 19, No. 4 pp. 197-202, 2018 https://doi.org/10.5762/kais.2018.19.4.197 ISSN 1975-4701 / eissn 2288-4688 핸드스캐닝작업방법에따라표준모델변형이작업정밀도에미치는정밀스캔에관한연구

More information

06_ÀÌÀçÈÆ¿Ü0926

06_ÀÌÀçÈÆ¿Ü0926 182 183 184 / 1) IT 2) 3) IT Video Cassette Recorder VCR Personal Video Recorder PVR VCR 4) 185 5) 6) 7) Cloud Computing 8) 186 VCR P P Torrent 9) avi wmv 10) VCR 187 VCR 11) 12) VCR 13) 14) 188 VTR %

More information

Comparison of digital models generated from three-dimensional optical scanner and cone beam computed tomography 에서시도되어왔고특히교정영역에서의활용가능성은여러연구에서확인되었다. 14

Comparison of digital models generated from three-dimensional optical scanner and cone beam computed tomography 에서시도되어왔고특히교정영역에서의활용가능성은여러연구에서확인되었다. 14 http://dx.doi.org/10.14368/jdras.2016.32.1.60 ISSN 2384-4353 eissn 2384-4272 Original Article Comparison of digital models generated from three-dimensional optical scanner and cone beam computed tomography

More information

012임수진

012임수진 Received : 2012. 11. 27 Reviewed : 2012. 12. 10 Accepted : 2012. 12. 12 A Clinical Study on Effect of Electro-acupuncture Treatment for Low Back Pain and Radicular Pain in Patients Diagnosed with Lumbar

More information

대한치과이식(임프란트)학회 학회지 제 27권 1호

대한치과이식(임프란트)학회 학회지 제 27권 1호 Hounsfield unit을 이용한 임프란트 식립부위의 골질 평가 조선대학교 치과대학 구강악안면외과학교실 문성용, 정미경, 김수관, 김학균 I. 서론 최근 임프란트 무치악의 재건을 위해 많이 사용되어 지고 있으며, 그 성공률은 환자의 전신적 건강 상태, 임프란트 재료의 생체 적합성, 술자의 술식, 골의 질 과 양에 따라 좌우된다 ). bone quantity"는

More information

<35335FBCDBC7D1C1A42DB8E2B8AEBDBAC5CDC0C720C0FCB1E2C0FB20C6AFBCBA20BAD0BCAE2E687770>

<35335FBCDBC7D1C1A42DB8E2B8AEBDBAC5CDC0C720C0FCB1E2C0FB20C6AFBCBA20BAD0BCAE2E687770> Journal of the Korea Academia-Industrial cooperation Society Vol. 15, No. 2 pp. 1051-1058, 2014 http://dx.doi.org/10.5762/kais.2014.15.2.1051 멤리스터의 전기적 특성 분석을 위한 PSPICE 회로 해석 김부강 1, 박호종 2, 박용수 3, 송한정 1*

More information

03이경미(237~248)ok

03이경미(237~248)ok The recent (2001-2010) changes on temperature and precipitation related to normals (1971-2000) in Korea* Kyoungmi Lee** Hee-Jeong Baek*** ChunHo Cho**** Won-Tae Kwon*****. 61 (1971~2000) 10 (2001~2010).

More information

KEIT PD Issue Report PD ISSUE REPORT FEBRUARY 2014 VOL 치과의료기기개요및특징 1.1 정의 치과용의료기기란치아와그주위조직및구강을포함한악안면영역의질병이나비정상적상태등을예방하고진단하며치료를도모하는의학의한분야인치과진료및

KEIT PD Issue Report PD ISSUE REPORT FEBRUARY 2014 VOL 치과의료기기개요및특징 1.1 정의 치과용의료기기란치아와그주위조직및구강을포함한악안면영역의질병이나비정상적상태등을예방하고진단하며치료를도모하는의학의한분야인치과진료및 치과의료기기의기술및산업동향 l 저자 l 허영 PD / KEIT 의료기기 PD 실양종수수석 / KEIT 의료기기 PD 실박경환선임 / KEIT 의료기기 PD 실신상완소장 / 고려대학교임상치의학연구소이정렬교수 / 고려대학교구로병원치과센터강태건박사 / 오송첨단의료기기개발지원센터 SUMMARY 치과의료기기첨단기술융합치과용진단영상장비, 치과인상채득, 보철물설계및가공분야의디지털화를통한효율성증대디지털정보교환을통한다양한치과첨단의료기기장비및재료의상호연관성에의한치과의료기기전반적성장기대치과의료기술의디지털화

More information

지능정보연구제 16 권제 1 호 2010 년 3 월 (pp.71~92),.,.,., Support Vector Machines,,., KOSPI200.,. * 지능정보연구제 16 권제 1 호 2010 년 3 월

지능정보연구제 16 권제 1 호 2010 년 3 월 (pp.71~92),.,.,., Support Vector Machines,,., KOSPI200.,. * 지능정보연구제 16 권제 1 호 2010 년 3 월 지능정보연구제 16 권제 1 호 2010 년 3 월 (pp.71~92),.,.,., Support Vector Machines,,., 2004 5 2009 12 KOSPI200.,. * 2009. 지능정보연구제 16 권제 1 호 2010 년 3 월 김선웅 안현철 社 1), 28 1, 2009, 4. 1. 지능정보연구제 16 권제 1 호 2010 년 3 월 Support

More information

,. 3D 2D 3D. 3D. 3D.. 3D 90. Ross. Ross [1]. T. Okino MTD(modified time difference) [2], Y. Matsumoto (motion parallax) [3]. [4], [5,6,7,8] D/3

,. 3D 2D 3D. 3D. 3D.. 3D 90. Ross. Ross [1]. T. Okino MTD(modified time difference) [2], Y. Matsumoto (motion parallax) [3]. [4], [5,6,7,8] D/3 Depth layer partition 2D 3D a), a) 3D conversion of 2D video using depth layer partition Sudong Kim a) and Jisang Yoo a) depth layer partition 2D 3D. 2D (depth map). (edge directional histogram). depth

More information

I 서론 치과용 임플란트는 Brånemark 등1의 골유착 (osseointegration) 발견 이후 끊임없는 발전 을 거듭해 왔다. Brånemark 등 1 이 밝혀낸 골 유착은 임플란트의 표면과 living bone 사이에 연조직 층 의 생성이 없이 직접 골조직이

I 서론 치과용 임플란트는 Brånemark 등1의 골유착 (osseointegration) 발견 이후 끊임없는 발전 을 거듭해 왔다. Brånemark 등 1 이 밝혀낸 골 유착은 임플란트의 표면과 living bone 사이에 연조직 층 의 생성이 없이 직접 골조직이 ISSN 1229-5418 ; 17(4): 268~274 임플란트 드릴링 시 발생하는 열과 그 원인에 대한 분석 한선희 1, 김병국 2, 김범익 2 1 서울대학교 치의학대학원 치주과학교실, 2 오스템임플란트연구소 Temperature Measurement during Implant-site Preparation Seon-Hee Han 1, Byoung Kook

More information

<313630313032C6AFC1FD28B1C7C7F5C1DF292E687770>

<313630313032C6AFC1FD28B1C7C7F5C1DF292E687770> 양성자가속기연구센터 양성자가속기 개발 및 운영현황 DOI: 10.3938/PhiT.25.001 권혁중 김한성 Development and Operational Status of the Proton Linear Accelerator at the KOMAC Hyeok-Jung KWON and Han-Sung KIM A 100-MeV proton linear accelerator

More information

저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할

저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할 저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할수없습니다. 변경금지. 귀하는이저작물을개작, 변형또는가공할수없습니다. 귀하는, 이저작물의재이용이나배포의경우,

More information

좋은 사진 찍는 방법

좋은 사진 찍는 방법 Based on Photo Zone by Klaus Schroiff (Klaus@photozone.de) Translation & Edit by Jihoon Jason Wang (DS2SJT / jasonw@korea.com) - Prologue.. And.. special thanks to Klaus Jason Jihoon Wang (jasonw@korea.com)

More information

untitled

untitled J Dent Hyg Sci Vol.14, No.2, 2014, pp.94-100 RESEARCH ARTICLE 치과용탄성인상재를기반으로채득된디지털모형의정확성연구 : 3 차원중첩분석 김기백ㆍ정재관 1 ㆍ김재홍 고려대학교일반대학원보건과학과치의기공전공, 1 대전보건대학교치기공학과 Accuracy of Digital Impression Made from Different

More information

386-390.hwp

386-390.hwp 386 HANYANG MEDICAL REVIEWS Vol. 29 No. 4, 2009 우리나라 미숙아의 통계와 의료비용 Statistics and Medical Cost of Preterm in Korea 윤혜선 을지대학교 노원을지병원 소아청소년과학교실 Hye Sun Yoon, M.D., Ph.D., Department of Pediatrics, Nowon

More information

CASE REPORT 시멘트유지형임플란트보철물의신개념유지방식 : Fiber post 의설측수평삽입에의한유지력보완 윤나리 이성복 * 이석원 안수진 박수정강동경희대학교치과병원생체재료보철과 A new

CASE REPORT   시멘트유지형임플란트보철물의신개념유지방식 : Fiber post 의설측수평삽입에의한유지력보완 윤나리 이성복 * 이석원 안수진 박수정강동경희대학교치과병원생체재료보철과 A new CASE REPORT https://doi.org/10.4047/jkap.2017.55.1.71 시멘트유지형임플란트보철물의신개념유지방식 : Fiber post 의설측수평삽입에의한유지력보완 윤나리 이성복 * 이석원 안수진 박수정강동경희대학교치과병원생체재료보철과 A new retaining method of cement-retained restoration with

More information

I

I I II III (C B ) (C L ) (HL) Min c ij x ij f i y i i H j H i H s.t. y i 1, k K, i W k C B C L p (HL) x ij y i, i H, k K i, j W k x ij y i {0,1}, i, j H. K W k k H K i i f i i d ij i j r ij i j c ij r ij

More information

À±½Â¿í Ãâ·Â

À±½Â¿í Ãâ·Â Representation, Encoding and Intermediate View Interpolation Methods for Multi-view Video Using Layered Depth Images The multi-view video is a collection of multiple videos, capturing the same scene at

More information

10(3)-09.fm

10(3)-09.fm w y wz 10«3y 253~258 (2010.12.) Journal of Korean Society of Urban Environment ³ w Á» Á Á y w y œw (2010 11 22, 2010 12 9 k) Study on Determine of Detention Pond in Small Developed Area In-Soo Chang ½

More information

<303220B1E8C7FDBFB52E666D>

<303220B1E8C7FDBFB52E666D> Journal of Dental Hygiene Science Vol. 11, No. 4, pp. 299~303 (2011) 전치부전부도재관을이용한테트라싸이클린착색의심미보철회복증례 (IPS e.max CAD 블록의이용 ) 김재홍 조영규 1 김혜영 고려대학교대학원보건과학과치의기공전공, 1 엘리시아치과의원 Esthetic Rehabilitation of Anterior

More information

Journal of Educational Innovation Research 2017, Vol. 27, No. 2, pp DOI: * Review of Research

Journal of Educational Innovation Research 2017, Vol. 27, No. 2, pp DOI:   * Review of Research Journal of Educational Innovation Research 2017, Vol. 27, No. 2, pp.79-102 DOI: http://dx.doi.org/10.21024/pnuedi.27.2.201706.79 * Review of Research Trends on Curriculum for Students with Severe and multiple

More information

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Mar.; 25(3),

THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE Mar.; 25(3), THE JOURNAL OF KOREAN INSTITUTE OF ELECTROMAGNETIC ENGINEERING AND SCIENCE. 2014 Mar.; 25(3), 304310. http://dx.doi.org/10.5515/kjkiees.2014.25.3.304 ISSN 1226-3133 (Print)ISSN 2288-226X (Online) Analysis

More information

CASE REPORT 전악무치악환자에서골이식술을선행한임플란트고정성보철수복증례 안주남 1 이정진 1 서재민 1 김경아 2 * 1 전북대학교치과대학치과보철학교실및구강생체과학연구소, 2 을지대학교의과

CASE REPORT   전악무치악환자에서골이식술을선행한임플란트고정성보철수복증례 안주남 1 이정진 1 서재민 1 김경아 2 * 1 전북대학교치과대학치과보철학교실및구강생체과학연구소, 2 을지대학교의과 CASE REPORT https://doi.org/10.4047/jkap.2018.56.1.77 전악무치악환자에서골이식술을선행한임플란트고정성보철수복증례 안주남 1 이정진 1 서재민 1 김경아 2 * 1 전북대학교치과대학치과보철학교실및구강생체과학연구소, 2 을지대학교의과대학치과학교실 Full mouth rehabilitation of fully edentulous

More information

ORIGINAL ARTICLE 구치부임플란트고정성수복물에서의지대주나사풀림현상과 이에영향을미치는요인 홍수정 배정윤 * 김현희 가천대학교길병원치과보철과 The incidence of the abu

ORIGINAL ARTICLE   구치부임플란트고정성수복물에서의지대주나사풀림현상과 이에영향을미치는요인 홍수정 배정윤 * 김현희 가천대학교길병원치과보철과 The incidence of the abu ORIGINAL ARTICLE https://doi.org/10.4047/jkap.2018.56.3.212 구치부임플란트고정성수복물에서의지대주나사풀림현상과 이에영향을미치는요인 홍수정 배정윤 * 김현희 가천대학교길병원치과보철과 The incidence of the abutment screw loosening and its affecting factors in

More information

01

01 SFI-Bar (Stress-Free on Implants) Innovative prefabricated Bar system for Implant (주 ) 성 원 덴 탈 www.sungwondental.co.kr 2-Implant the new ingenious stress-free bar system on implants. Stress-free, excellent

More information

7-2È£

7-2È£ original article OA A comparative study on accuracy of the implant torque devices for screw tightening Hi-Chan Jung 1, Chang-Mo Jeong 2 1 Graduate student, DDS,MSD Dept of Prosthodontic Dentistry, College

More information

00)13-3목차1~3

00)13-3목차1~3 An evaluation of marginal bone level change in single short implant : an 1-year follow up Tae-Hoon Lee, Joo-Hyun Kwon, Sunjai Kim, Chong-Hyun Han Department of Prosthodontics, Kangnam Severance hospital,

More information

clinical case of hybrid telescopic double crown using friction pin with an isolated few remaining teeth 기에 pin을삽입하여내관과 pin의활성화에의하여유지를조절할수있는특징을가지고있다. 6

clinical case of hybrid telescopic double crown using friction pin with an isolated few remaining teeth 기에 pin을삽입하여내관과 pin의활성화에의하여유지를조절할수있는특징을가지고있다. 6 http://dx.doi.org/10.14368/jdras.2016.32.1.80 ISSN 2384-4353 eissn 2384-4272 Case Report clinical case of hybrid telescopic double crown using friction pin with an isolated few remaining teeth Sang-Hyuk

More information

3 : 3D (Seunggi Kim et. al.: 3D Depth Estimation by a Single Camera) (Regular Paper) 24 2, (JBE Vol. 24, No. 2, March 2019)

3 : 3D (Seunggi Kim et. al.: 3D Depth Estimation by a Single Camera) (Regular Paper) 24 2, (JBE Vol. 24, No. 2, March 2019) 3 : 3D (Regular Paper) 24 2, 2019 3 (JBE Vol. 24, No. 2, March 2019) https://doi.org/10.5909/jbe.2019.24.2.281 ISSN 2287-9137 (Online) ISSN 1226-7953 (Print) 3D a), a), a), a) 3D Depth Estimation by a

More information

A 617

A 617 Special Issue Diabetic Retinopathy Won Ki Lee, M.D. Department of Ophthalmology The Catholic University of Korea College of Medicine Kangnam St. Mary s Hospital E mail : wklee@catholic.ac.kr Abstract R

More information

Print

Print 22-12324-4TEL:3668-3114 FAX:742-3387 TEL:3668-3120 FAX:745-9476 TEL:3668-3109, 2279-0867~8 TEL:3668-3127 TEL:3668-3123, 3128, 3162 www.saeki.co.kr, www.pentaximaging.co.kr Small 의 큰 스타일을 경험하다 당신의 카메라만으로도,

More information

본증례는하악은 locator attachment 오버덴처를사용했으나, 임플란트의식립각도가좋지못했고임플란트의숫자도부족하여임플란트가실패한상태였다. 상악은심한치주염이동반된소수잔존치상태였던환자를 hybrid telescopic double crown with friction

본증례는하악은 locator attachment 오버덴처를사용했으나, 임플란트의식립각도가좋지못했고임플란트의숫자도부족하여임플란트가실패한상태였다. 상악은심한치주염이동반된소수잔존치상태였던환자를 hybrid telescopic double crown with friction CASE REPORT http://dx.doi.org/10.4047/jkap.2016.54.3.246 상악의소수잔존치와실패한하악의임플란트오버덴처의재수복임상증례 : 프릭션핀을이용한하이브리드텔레스코픽이중관법 하석준 조진현 * 경북대학교치과대학치과보철학교실 Hybrid telescopic double crown denture on maxillary few remaining

More information

16(1)-3(국문)(p.40-45).fm

16(1)-3(국문)(p.40-45).fm w wz 16«1y Kor. J. Clin. Pharm., Vol. 16, No. 1. 2006 x w$btf3fqpsu'psn û w m w Department of Statistics, Chonnam National University Eunsik Park College of Natural Sciences, Chonnam National University

More information

(JBE Vol. 23, No. 5, September 2018) (Regular Paper) 23 5, (JBE Vol. 23, No. 5, September 2018) ISSN

(JBE Vol. 23, No. 5, September 2018) (Regular Paper) 23 5, (JBE Vol. 23, No. 5, September 2018)   ISSN (JBE Vol. 23, No. 5, September 2018) (Regular Paper) 23 5, 2018 9 (JBE Vol. 23, No. 5, September 2018) https://doi.org/10.5909/jbe.2018.23.5.636 ISSN 2287-9137 (Online) ISSN 1226-7953 (Print) a), a) The

More information

김범수

김범수 Analysis of Outcomes after Resection of Sarcomatous Hepatocellular Carcinoma Purpose: Sarcomatous hepatocellular carcinoma (HCC) is rare. Therefore, the clinicopathologic characteristics and prognosis

More information

디지털임플란트시술계획및수술가이드를선택해야하는 5 가지이유 Implant Studio 를사용해야하는이유 환자의진료시간및방문횟수감소 디지털정확도활용및수동오류감소 보철물기반시술계획 개방형시스템플랫폼 유연한디지털워크플로 경제적인비용으로수술가이드를로컬로제작 유연한도구및기능 최종

디지털임플란트시술계획및수술가이드를선택해야하는 5 가지이유 Implant Studio 를사용해야하는이유 환자의진료시간및방문횟수감소 디지털정확도활용및수동오류감소 보철물기반시술계획 개방형시스템플랫폼 유연한디지털워크플로 경제적인비용으로수술가이드를로컬로제작 유연한도구및기능 최종 Implant Studio 임플란트시술계획및수술가이드디자인 디지털임플란트시술계획및수술가이드를선택해야하는 5 가지이유 Implant Studio 를사용해야하는이유 환자의진료시간및방문횟수감소 디지털정확도활용및수동오류감소 보철물기반시술계획 개방형시스템플랫폼 유연한디지털워크플로 경제적인비용으로수술가이드를로컬로제작 유연한도구및기능 최종임플란트위치와복원결과에대한예측가능성향상

More information

저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할

저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할 저작자표시 - 비영리 - 변경금지 2.0 대한민국 이용자는아래의조건을따르는경우에한하여자유롭게 이저작물을복제, 배포, 전송, 전시, 공연및방송할수있습니다. 다음과같은조건을따라야합니다 : 저작자표시. 귀하는원저작자를표시하여야합니다. 비영리. 귀하는이저작물을영리목적으로이용할수없습니다. 변경금지. 귀하는이저작물을개작, 변형또는가공할수없습니다. 귀하는, 이저작물의재이용이나배포의경우,

More information

DBPIA-NURIMEDIA

DBPIA-NURIMEDIA 박건수 *, 서태영 **, 김종욱 *** ". 요약 Abstract The induction melting furnace using electric generator has been introduced since 1920s, and it began to be widely applied to industrial applications due to increasing

More information

CASE REPORT 일정한간극으로새롭게고안된완압형볼어태치먼트를이용한피개의치수복 유진주 김만용 윤준호 * 국민건강보험일산병원치과보철과 Overdentures using newly designe

CASE REPORT   일정한간극으로새롭게고안된완압형볼어태치먼트를이용한피개의치수복 유진주 김만용 윤준호 * 국민건강보험일산병원치과보철과 Overdentures using newly designe CASE REPORT https://doi.org/10.4047/jkap.2017.55.3.311 유진주 김만용 윤준호 * 국민건강보험일산병원치과보철과 Overdentures using newly designed metal ball attachment containing predetermined gap with stress breakers Jin-Joo Yoo,

More information